MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-04-24 for SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC manufactured by Philips Ultrasound, Inc.
[143043546]
Evaluation of the power module will be included in a follow up report upon its return and investigation completion.
Patient Sequence No: 1, Text Type: N, H10
[143043547]
A customer reported their affiniti ultrasound system had produced a burning odor and a small flame from the back of the unit. No patient or user was harmed as a result of the issue. The site? S local service engineer inspected the system and determined the problem was caused by a failed power module. After replacing the power module and power cord, the system passed all tests and was returned to service.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 3019216-2019-00019 |
MDR Report Key | 8547115 |
Date Received | 2019-04-24 |
Date of Report | 2019-03-04 |
Date Mfgr Received | 2019-03-04 |
Date Added to Maude | 2019-04-24 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MR. PAUL CORRIGAN |
Manufacturer Street | 22100 BOTHELL EVERETT HIGHWAY |
Manufacturer City | BOTHELL WA 98021 |
Manufacturer Country | US |
Manufacturer Postal | 98021 |
Manufacturer Phone | 4254877000 |
Manufacturer G1 | PHILIPS MEDICAL SYSTEMS |
Manufacturer Street | 3000 MINUTEMAN ROAD |
Manufacturer City | ANDOVER MA 01810 |
Manufacturer Country | US |
Manufacturer Postal Code | 01810 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 0 |
Generic Name | SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC |
Product Code | IYN |
Date Received | 2019-04-24 |
Returned To Mfg | 2019-04-24 |
Lot Number | US815F1283 |
Operator | HEALTH PROFESSIONAL |
Device Availability | R |
Device Age | DA |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | PHILIPS ULTRASOUND, INC |
Manufacturer Address | 22100 BOTHELL EVERETT HIGHWAY BOTHELL WA 98021 US 98021 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2019-04-24 |