CHEMSTRIP MICRAL URINE TEST STRIPS

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-04-25 for CHEMSTRIP MICRAL URINE TEST STRIPS manufactured by Roche Diagnostics.

Event Text Entries

[143417023] The customer's product was returned for investigation. The test strips did not show any abnormalities. The customer material and relevant retention material were tested with five different human serum albumin solutions and the results from the investigation were in accordance with the testing plan. No product failure was detected. There were no follow up/corrective actions for this event. The device was returned. This device is not labeled for single use and is not reprocessed or reused.
Patient Sequence No: 1, Text Type: N, H10


[143417024] This report summarizes 1 malfunction event. Low results were generated by the chemstrip micral. The event involved a total of 1 patient. The patient's age is (b)(6). The patient is male.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2019-90131
MDR Report Key8549728
Date Received2019-04-25
Date Added to Maude2019-04-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA0
Report to FDA3
Event Location3
Manufacturer ContactMICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG), NA 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameCHEMSTRIP MICRAL URINE TEST STRIPS
Generic NameURINARY PROTEIN OR ALBUMIN (NONQUANTITATIVE) TEST SYSTEM
Product CodeJIR
Date Received2019-04-25
Model NumberMICRAL
Lot Number33829802
Device Availability*
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 46250 US 46250


Patients

Patient NumberTreatmentOutcomeDate
10 2019-04-25

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.