ELECSYS HSV-1 IGG IMMUNOASSAY 05572185190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2019-04-25 for ELECSYS HSV-1 IGG IMMUNOASSAY 05572185190 manufactured by Roche Diagnostics.

Event Text Entries

[143413945] The investigations for the 1 event did not identify a product problem. The cause of the event could not be determined. Non-reactive test results do not completely rule out the possibility of hsv-1 infection. Individuals may not exhibit any detectable igg antibodies at an early stage of acute infection. False negative results may occur when the hs-virus is glycoprotein g (gg) deficient. The follow up/corrective actions for the 1 event was as follows: the patient sample was requested for investigation but could not be provided. No devices were returned. This device is not labeled for single use and is not reprocessed or reused.
Patient Sequence No: 1, Text Type: N, H10


[143413946] This report summarizes 1 malfunction event. An erroneous low result was generated by the cobas e 411 immunoassay analyzer. The events involved a total of 1 patient with the following: a low elecsys hsv-1 igg immunoassay (hsv-1 igg) result.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2019-90090
MDR Report Key8549742
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2019-04-25
Date of Report2019-04-25
Date of Event2019-03-20
Date Mfgr Received2019-03-25
Date Added to Maude2019-04-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG), NA 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameELECSYS HSV-1 IGG IMMUNOASSAY
Generic NameHERPES SIMPLEX VIRUS, HSV-1
Product CodeMXJ
Date Received2019-04-25
Model NumberHSV-1
Catalog Number05572185190
Lot Number30828200
Device Expiration Date2019-04-30
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 46250 US 46250


Patients

Patient NumberTreatmentOutcomeDate
10 2019-04-25