CRPL3 C-REACTIVE PROTEIN GEN.3 CRP G3

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-04-25 for CRPL3 C-REACTIVE PROTEIN GEN.3 CRP G3 manufactured by Roche Diagnostics.

Event Text Entries

[143416410] The investigations for the 1 event did not identify a product problem. The cause of the event could not be determined. The samples were submitted for investigation where the customer's high crpl3 results were not confirmed. The customer's repeat results were confirmed during the investigation. The follow up/corrective actions for the 1 event were as follows: the field service engineer adjusted the gear pump pressure, checked the rinsing of the cuvettes which were ok and replaced reagent and sample probes. Qc was acceptable. Sample material was requested for investigation. No devices were returned. This device is not labeled for single use and is not reprocessed or reused.
Patient Sequence No: 1, Text Type: N, H10


[143416411] This report summarizes 1 malfunction event. An erroneous low result was generated by the cobas e 411 immunoassay analyzer. The event involved a total of 3 patients with high crpl3 c-reactive protein gen. 3 results.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2019-90091
MDR Report Key8549746
Date Received2019-04-25
Date Added to Maude2019-04-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA0
Report to FDA3
Event Location3
Manufacturer ContactMICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG), NA 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameCRPL3 C-REACTIVE PROTEIN GEN.3
Generic NameC-REACTIVE PROTEIN IMMUNOLOGICAL TEST SYSTEM
Product CodeDCN
Date Received2019-04-25
Model NumberCRP G3
Lot Number383790
Device Availability*
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 46250 US 46250


Patients

Patient NumberTreatmentOutcomeDate
10 2019-04-25

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.