TOSOH HLC-723G8 ANALYZER G8 021560

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2019-04-25 for TOSOH HLC-723G8 ANALYZER G8 021560 manufactured by Tosoh Corporation.

Event Text Entries

[146531957] For all 7 reported events, a field service engineer (fse) evaluated the analyzers at the customer sites and was able to confirm the pump unit errors. The fse replaced the pump components and assemblies for 6 of the 7 events. And for 1 of 7 events, the fse found air in lines which was flushed out. 2 of 7 pumps were returned to instrument service center, and further confirmed pumps were faulty; 1 had a cam shaft fault and the other had faulty check valves. No additional issues have been reported by the facilities. All seven (7) g8 analyzers were repaired and returned to operational status.
Patient Sequence No: 1, Text Type: N, H10


[146531958] This report summarizes 7 malfunction events for the g8 analyzers. The review of events indicated that the g8 analyzer experienced pump error messages, which caused delay in reporting of patient results for hemoglobin a1c (hba1c). These reports were received from various sources. There was no indication of patient intervention or adverse health consequences due to the delayed reporting in patient results. None of these events necessitated remedial action to prevent an unreasonable risk of substantial harm to the public health.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number8031673-2019-00111
MDR Report Key8550797
Report SourceHEALTH PROFESSIONAL
Date Received2019-04-25
Date of Report2019-08-09
Date Added to Maude2019-04-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA0
Report to FDA3
Event Location3
Manufacturer ContactMS. DORIA ESQUIVEL
Manufacturer StreetSHIBA-KOEN FIRST BUILDING 3-8-2 SHIBA
Manufacturer CityMINATO-KU, TOKYO, JAPAN 1058623
Manufacturer CountryJA
Manufacturer Postal1058623
Manufacturer G1TOSOH CORPORATION
Manufacturer StreetSHIBA-KOEN FIRST BUILDING 3-8-2 SHIBA
Manufacturer CityMINATO-KU, TOKYO, JAPAN 1058623
Manufacturer CountryJA
Manufacturer Postal Code1058623
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTOSOH HLC-723G8 ANALYZER G8
Generic NameG8
Product CodeLCP
Date Received2019-04-25
Returned To Mfg2019-04-12
Model NumberG8
Catalog Number021560
Device AvailabilityR
Device AgeDA
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerTOSOH CORPORATION
Manufacturer AddressSHIBA-KOEN FIRST BUILDING 3-8-2 SHIBA MINATO-KU, TOKYO, JAPAN 1058623 JA 1058623


Patients

Patient NumberTreatmentOutcomeDate
10 2019-04-25

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.