[143239250] .
Patient Sequence No: 1, Text Type: N, H10


[143239251] As reported by the prescribing / fitting practitioner ((b)(6)), "we recently had a patient being fit with a p4 pump. During the test fit phase, our technicians had scratched up the top of the mounting plate in order to give it more bonding surface area. The mounting plate was attached to the socket using fab tec two part resin. Then the plate and socket were bonded together with fiberglass wrap. The patient left our facility and later that night the mounting plate completely separated from the socket. Unfortunately, the patient fell and dislocated his shoulder. "
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1721652-2019-00002
MDR Report Key8551262
Date Received2019-04-25
Date of Report2019-04-25
Date of Event2019-03-12
Date Mfgr Received2019-03-29
Device Manufacturer Date2019-02-28
Date Added to Maude2019-04-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. STEPHEN ANDERSON
Manufacturer Street3820 WEST GREAT LAKES DRIVE
Manufacturer CityWEST VALLEY CITY UT 84120
Manufacturer CountryUS
Manufacturer Postal84120
Manufacturer Phone8019746676
Manufacturer G1OTTO BOCK HEALTHCARE LP
Manufacturer Street3820 WEST GREAT LAKES DRIVE
Manufacturer CityWEST VALLEY CITY UT 84120
Manufacturer CountryUS
Manufacturer Postal Code84120
Single Use3
Remedial ActionOT
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameOTTOBOCK HARMONY? P4 HD MECHANICAL PUMP
Generic Name4R181 HARMONY? P4 HD MECHANICAL PUMP
Product CodeISP
Date Received2019-04-25
Model Number4R181
Catalog Number4R181
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerOTTO BOCK HEALTHCARE LP
Manufacturer Address3820 WEST GREAT LAKES DRIVE WEST VALLEY CITY UT 84120 US 84120


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2019-04-25