MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2019-04-25 for TOSOH HLC-723G8 ANALYZER G8 021560 manufactured by Tosoh Corporation.
[146532044]
A field service engineer (fse) evaluated the reported event at the customer's site. The fse was not able to confirm or replicate the reported event. The fse suspected that after the customer installed a new column, the g8 analyzer was not properly primed, which caused the calibration to fail.
Patient Sequence No: 1, Text Type: N, H10
[146532045]
This report summarizes 1 malfunction event on the g8 analyzer. The review of this event indicated that the g8 analyzer experienced a calibration failure, which caused delayed reporting of hemoglobin a1c (hba1c) patient results. This report was received from one (1) source. There was no indication of patient intervention or adverse health consequences due to the delayed reporting in patient results. This event did not necessitate remedial action to prevent an unreasonable risk of substantial harm to the public health.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 8031673-2019-00055 |
MDR Report Key | 8552773 |
Report Source | HEALTH PROFESSIONAL |
Date Received | 2019-04-25 |
Date of Report | 2019-08-02 |
Date Mfgr Received | 2019-07-03 |
Date Added to Maude | 2019-04-25 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 0 |
Initial Report to FDA | 0 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MS. DORIA ESQUIVEL |
Manufacturer Street | SHIBA-KOEN FIRST BUILDING 3-8-2 SHIBA |
Manufacturer City | MINATO-KU, TOKYO 1058623 |
Manufacturer Country | JA |
Manufacturer Postal | 1058623 |
Manufacturer G1 | TOSOH CORPORATION |
Manufacturer Street | SHIBA-KOEN FIRST BUILDING 3-8-2 SHIBA |
Manufacturer City | MINATO-KU, TOKYO 1058623 |
Manufacturer Country | JA |
Manufacturer Postal Code | 1058623 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | TOSOH HLC-723G8 ANALYZER G8 |
Generic Name | G8 |
Product Code | LCP |
Date Received | 2019-04-25 |
Model Number | G8 |
Catalog Number | 021560 |
Device Availability | Y |
Device Age | DA |
Device Eval'ed by Mfgr | * |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | TOSOH CORPORATION |
Manufacturer Address | SHIBA-KOEN FIRST BUILDING 3-8-2 SHIBA MINATO-KU, TOKYO 1058623 JA 1058623 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2019-04-25 |