MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2019-04-25 for TOSOH HLC-723G8 ANALYZER G8 021560 manufactured by Tosoh Corporation.
[146474302]
Field service engineers (fse) evaluated 8 of the 9 injection valves on the g8 analyzers at the customers' sites. One event was evaluated by the fse over-the-phone with the customer. The fses were able to confirm the reported events. The 9 g8 analyzers were repaired by either adjustments, cleaning, or replacing the injection valves, and returned to operational status. Four (4) of the 7 injection valves were returned for investigation to the tosoh instrument service center (isc). The investigation by isc replicated the reported events on 2 of the 4 returned injection valves.
Patient Sequence No: 1, Text Type: N, H10
[146474303]
This report summarizes
Patient Sequence No: 1, Text Type: D, B5
| Report Number | 8031673-2019-00124 |
| MDR Report Key | 8552818 |
| Report Source | HEALTH PROFESSIONAL |
| Date Received | 2019-04-25 |
| Date of Report | 2019-07-02 |
| Date Mfgr Received | 2019-06-28 |
| Date Added to Maude | 2019-04-25 |
| Event Key | 0 |
| Report Source Code | Manufacturer report |
| Manufacturer Link | Y |
| Number of Patients in Event | 0 |
| Adverse Event Flag | 3 |
| Product Problem Flag | 3 |
| Reprocessed and Reused Flag | 3 |
| Health Professional | 0 |
| Initial Report to FDA | 0 |
| Report to FDA | 3 |
| Event Location | 3 |
| Manufacturer Contact | MS. DORIA ESQUIVEL |
| Manufacturer Street | SHIBA-KOEN FIRST BUILDING 3-8-2 SHIBA |
| Manufacturer City | MINATO-KU, TOKYO 1058623 |
| Manufacturer Country | JA |
| Manufacturer Postal | 1058623 |
| Manufacturer G1 | TOSOH CORPORATION |
| Manufacturer Street | SHIBA-KOEN FIRST BUILDING 3-8-2 SHIBA |
| Manufacturer City | MINATO-KU, TOKYO 1058623 |
| Manufacturer Country | JA |
| Manufacturer Postal Code | 1058623 |
| Single Use | 3 |
| Previous Use Code | 3 |
| Event Type | 3 |
| Type of Report | 3 |
| Brand Name | TOSOH HLC-723G8 ANALYZER G8 |
| Generic Name | G8 |
| Product Code | LCP |
| Date Received | 2019-04-25 |
| Model Number | G8 |
| Catalog Number | 021560 |
| Device Availability | Y |
| Device Age | DA |
| Device Eval'ed by Mfgr | * |
| Device Sequence No | 1 |
| Device Event Key | 0 |
| Manufacturer | TOSOH CORPORATION |
| Manufacturer Address | SHIBA-KOEN FIRST BUILDING 3-8-2 SHIBA MINATO-KU, TOKYO 1058623 JA 1058623 |
| Patient Number | Treatment | Outcome | Date |
|---|---|---|---|
| 1 | 0 | 2019-04-25 |