MAXIMA ELITE 1:5 6361849

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2019-04-26 for MAXIMA ELITE 1:5 6361849 manufactured by Sirona Dental Systems Gmbh.

Event Text Entries

[143259510] While no serious injury resulted in this event, there has been a previous report received where this malfunction resulted in a serious injury. Therefore, it must be presumed that recurrence of this malfunction could possibly cause or contribute to a serious injury or require medical or surgical intervention to preclude such. As such, this event is reportable per 21cfr part 803. The customer returned three handpieces because they were unsure which handpiece was involved in the event. The investigation has shown, that there was no manufacturing or material failure at all 3 instruments. The heating up could not be reproduced during the investigation. Therefore a heating up can only be reasoned due to an unintended use of the device, e. G. While pressing the push-button during the operation. This causes a contact between the rotor and the push button cover and finally results in destruction of the push button cover etc. And overheating.
Patient Sequence No: 1, Text Type: N, H10


[143259511] It was reported that a maxima elite 1:5 ca overheated and burned a patient. The patient was prescribed pain medication.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9614977-2019-00057
MDR Report Key8555620
Report SourceHEALTH PROFESSIONAL
Date Received2019-04-26
Date of Report2019-04-20
Date of Event2019-02-14
Date Added to Maude2019-04-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationDENTIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. KARL NITTINGER
Manufacturer Street221 W. PHILADELPHIA ST. SUITE 60W
Manufacturer CityYORK PA 17401
Manufacturer CountryUS
Manufacturer Postal17401
Manufacturer Phone7178494424
Manufacturer G1SIRONA DENTAL SYSTEMS GMBH
Manufacturer StreetFABRIKSTRASSE 31
Manufacturer CityBENSHEIM, D-64625
Manufacturer CountryGM
Manufacturer Postal CodeD-64625
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMAXIMA ELITE 1:5
Generic NameHANDPIECE, BELT AND/OR GEAR DRIVEN, DENTAL
Product CodeEFA
Date Received2019-04-26
Catalog Number6361849
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSIRONA DENTAL SYSTEMS GMBH
Manufacturer AddressFABRIKSTRASSE 31 BENSHEIM, D-64625 GM D-64625


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2019-04-26

© 2026 FDA.report