TC BARIATRIC PLUS W/AIR & PULM P1840RE300

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-04-26 for TC BARIATRIC PLUS W/AIR & PULM P1840RE300 manufactured by Hill-rom Batesville.

Event Text Entries

[143981614] The hill-rom technician found the patient left hand side rail needed to be repaired. The totalcare bariatric bed and totalcare bariatric plus therapy system require an effective maintenance program. We recommend that you perform a semi-annual preventative maintenance. Preventative maintenance will minimize downtime due to excessive wear. Test the bed function controls for proper operation of the function and momentary operation. Test all lockout controls and individually check for proper operation. The technician will repair the left hand side rail to resolve the issue. Based on this information, no further action is required.
Patient Sequence No: 1, Text Type: N, H10


[143981615] Hill-rom received a report from the account stating the head section would self run up. The bed was located at the account. There was no patient/user injury reported. This report was filed in our complaint handling system as complaint #(b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1824206-2019-00179
MDR Report Key8556855
Date Received2019-04-26
Date of Report2019-04-15
Date of Event2019-04-15
Date Mfgr Received2019-04-15
Device Manufacturer Date2006-11-09
Date Added to Maude2019-04-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactTONY WERNER
Manufacturer Street1069 STATE ROUTE 46 EAST
Manufacturer CityBATESVILLE IN 47006
Manufacturer CountryUS
Manufacturer Postal47006
Manufacturer Phone8129312359
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameTC BARIATRIC PLUS W/AIR & PULM
Generic NameBARIATRIC BED
Product CodeOSI
Date Received2019-04-26
Model NumberP1840RE300
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerHILL-ROM BATESVILLE
Manufacturer Address1069 STATE ROUTE 46 EAST BATESVILLE IN 47006 US 47006


Patients

Patient NumberTreatmentOutcomeDate
10 2019-04-26