ADVANTAGE PLUS

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2019-04-26 for ADVANTAGE PLUS manufactured by Medivators.

Event Text Entries

[143389304] A facility reported that they had not performed the waterline disinfection (wld) cycle in their advantage plus automated endoscope reprocessor (aer) in accordance with the aer user manual, causing potential risk of contamination of endoscopes reprocessed in the aer. Medivators regulatory followed up with the facility and informed them of the wld requirements per the aer user manual. The facility reported the internal 0. 1 micron absolute bacterial retentive filter was replaced every 6 months as required. However, the facility failed to perform the wld cycle after replacing the filter. The number of endoscopes reprocessed during the 18-month timeframe is unknown. There have been no reports of patient harm. This complaint will continue being monitored in medivators complaint handling system.
Patient Sequence No: 1, Text Type: N, H10


[143389305] A facility reported that they had not performed the waterline disinfection (wld) cycle in their advantage plus automated endoscope reprocessor (aer) in accordance with the aer user manual, causing potential risk of contamination of endoscopes reprocessed in the aer.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2150060-2019-00031
MDR Report Key8556946
Report SourceUSER FACILITY
Date Received2019-04-26
Date of Report2019-04-26
Date of Event2019-04-03
Date Mfgr Received2019-04-03
Device Manufacturer Date2011-12-01
Date Added to Maude2019-04-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactALEX NELSON
Manufacturer Street14605 28TH AVE N
Manufacturer CityMINNEAPOLIS MN 55447
Manufacturer CountryUS
Manufacturer Postal55447
Manufacturer Phone7635094799
Manufacturer G1MEDIVATORS
Manufacturer Street14605 28TH AVE N
Manufacturer CityMINNEAPOLIS MN 55447
Manufacturer CountryUS
Manufacturer Postal Code55447
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameADVANTAGE PLUS
Generic NameAUTOMATED ENDOSCOPE REPROCESSOR
Product CodeFEB
Date Received2019-04-26
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerMEDIVATORS
Manufacturer Address14605 28TH AVE N MINNEAPOLIS MN 55447 US 55447


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2019-04-26