CAPTURE-R READY INDICATOR RED CELLS 0006428

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2019-04-26 for CAPTURE-R READY INDICATOR RED CELLS 0006428 manufactured by Immucor, Inc..

Event Text Entries

[145778068] On april 15, 2019 immucor technical support used a remote electronic connection method to assess files on galileo echo v2 instrument serial number (b)(4); the camera report was acceptable and the event log indicated no errors at time of testing. Unable to rule out capture-r ready indicator red cells lot number 221291 as testing done april 15, 2019 used a different lot cric than the testing done april 3, 2019; however, the anti-e is showing dosage as heterozygous e cells are not reacting. On april 16, 2019 an immucor field service technician inspected galileo echo v2 instrument serial number (b)(4) at the customer facility. The technician found the instrument operational and performed a successful unexpected reaction checklist. Found instrument to be fully operational and to be operating in specification. The immucor internal report record number for this report is (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[145778069] On (b)(6) 2019 a customer reported an unexpected negative result for a screen assay on a galileo echo v2 instrument.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1034569-2019-00088
MDR Report Key8558105
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2019-04-26
Date of Report2019-04-26
Date of Event2019-04-03
Date Mfgr Received2019-04-15
Date Added to Maude2019-04-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR HOWARD YOREK
Manufacturer Street3130 GATEWAY DRIVE
Manufacturer CityNORCROSS GA 300915625
Manufacturer CountryUS
Manufacturer Postal300915625
Manufacturer Phone7704412051
Manufacturer G1IMMUCOR, INC.
Manufacturer Street3130 GATEWAY DRIVE
Manufacturer CityNORCROSS GA 300915625
Manufacturer CountryUS
Manufacturer Postal Code300915625
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCAPTURE-R READY INDICATOR RED CELLS
Generic NameQUALITY CONTROL KITS FOR BLOOD BANKING REAGENTS
Product CodeKSF
Date Received2019-04-26
Catalog Number0006428
Lot Number221291
Device Expiration Date2019-04-14
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerIMMUCOR, INC.
Manufacturer Address3130 GATEWAY DRIVE NORCROSS GA 300915625 US 300915625


Patients

Patient NumberTreatmentOutcomeDate
10 2019-04-26

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.