AXIOM LUMINOS TF 10093902

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign report with the FDA on 2019-04-29 for AXIOM LUMINOS TF 10093902 manufactured by Siemens Healthcare Gmbh, Advanced Therapies.

Event Text Entries

[143422268] The log files from the reported incident were deleted and the engineer has been unable to duplicate the situation. The investigation of the reported issue is on-going. A supplemental report will submitted if additional information becomes available. This report was submitted april 26, 2017.
Patient Sequence No: 1, Text Type: N, H10


[143422269] It was reported that during an exam the axiom luminos tf system started beeping fast and intermittently. An unintended system movement was reported as well. The system tilted on its own for several seconds before stopping. The operator waited for this behaviour to end and then completed the exam without further incidents. There are no injuries attributed to this event. According to siemens engineer, who inspected the system, no one was touching the system controls during the incident. The customer was advised to push the emergency stop bottom to stop all system movements in case of another incident.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3004977335-2017-76899
MDR Report Key8560202
Report SourceFOREIGN
Date Received2019-04-29
Date of Report2017-04-07
Date of Event2017-04-07
Date Mfgr Received2017-04-07
Date Added to Maude2019-04-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMEREDITH ADAMS
Manufacturer Street40 LIBERTY BLVD. 65-1A
Manufacturer CityMALVERN PA 19355
Manufacturer CountryUS
Manufacturer Postal19355
Manufacturer Phone6104486461
Manufacturer G1SIEMENS HEALTHCARE GMBH, ADVANCED THERAPIES
Manufacturer StreetSIEMENSSTRASSE 1
Manufacturer CityFORCHHEIM, GERMANY 91301
Manufacturer CountryGM
Manufacturer Postal Code91301
Single Use3
Remedial ActionIN
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameAXIOM LUMINOS TF
Generic NameSYSTEM, X-RAY, FLUOROSCOPIC, IMAGE-INTENSIFIED
Product CodeJAA
Date Received2019-04-29
Model Number10093902
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE GMBH, ADVANCED THERAPIES
Manufacturer AddressSIEMENSSTRASSE 1 FORCHHEIM, GERMANY 91301 GM 91301


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2019-04-29

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.