EN SNARE EN2007030/A

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,user f report with the FDA on 2019-04-29 for EN SNARE EN2007030/A manufactured by Merit Medical Ireland Ltd.

Event Text Entries

[143417523] The suspect device has been returned for evaluation. The complaint is confirmed. No definitive root cause could be determined however, it is likely that significant force was applied to the device during use. A search of the complaint database was performed and no similar complaints for this lot number were found. The device history record was reviewed and no exception documents were found.
Patient Sequence No: 1, Text Type: N, H10


[143417524] The account alleges that during an ivc filter removal procedure, the head of the vascular snare device detached. The clinician had successfully acquired antegrade venous access and during snare device manipulations, the head of the snare detached.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9616662-2019-00005
MDR Report Key8560892
Report SourceCOMPANY REPRESENTATIVE,USER F
Date Received2019-04-29
Date of Report2019-02-01
Date of Event2019-02-01
Date Mfgr Received2019-04-01
Device Manufacturer Date2018-12-14
Date Added to Maude2019-04-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. KATIE SWENSON CQE,CBA,CQPA.
Manufacturer Street1600 MERIT PARKWAY
Manufacturer CitySOUTH JORDAN UT 84095
Manufacturer CountryUS
Manufacturer Postal84095
Manufacturer Phone8012531600
Manufacturer G1MERIT MEDICAL IRELAND LTD
Manufacturer StreetPARKMORE BUSINESS PARK WEST BALLYBRIT
Manufacturer CityGALWAY,
Manufacturer CountryEI
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameEN SNARE
Generic NameENDOVASCULAR SNARE SYSTEM
Product CodeMMX
Date Received2019-04-29
Returned To Mfg2019-02-12
Catalog NumberEN2007030/A
Lot NumberK1490574
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMERIT MEDICAL IRELAND LTD
Manufacturer AddressPARKMORE BUSINESS PARK WEST BALLYBRIT GALWAY, EI


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2019-04-29