STERILCONTAINER S STER LOCKS PINK US910

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2019-04-29 for STERILCONTAINER S STER LOCKS PINK US910 manufactured by Aesculap Inc..

Event Text Entries

[143725609] Manufacturing site evaluation: investigation on-going. Additional information / investigation results will be provided in a supplemental report.
Patient Sequence No: 1, Text Type: N, H10


[143725610] It was reported that there was an issue with the locks; they were not changing color correctly after sterilization of the container. Prior to surgery, locks on an instrument set were noted as having turned to white instead of blue. There was a short delay (15 minutes) while sterilization cycles and monitoring records were checked to validate sterility. There was no patient harm. The central sterile processing (csp) staff confirmed that new locks were stored in a black bag to avoid light contamination. Additional information was not provided.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2916714-2019-00016
MDR Report Key8562635
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2019-04-29
Date of Report2019-04-29
Date of Event2019-04-03
Date Facility Aware2019-04-03
Date Mfgr Received2019-04-03
Date Added to Maude2019-04-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. NICOLE BROYLES
Manufacturer Street615 LAMBERT POINTE DRIVE
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer Phone3145515988
Manufacturer G1AESCULAP INC.
Manufacturer Street615 LAMBERT POINTE DRIVE
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal Code63042
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSTERILCONTAINER S STER LOCKS PINK
Generic NameSTERILE TECHNOLOGY
Product CodeFRG
Date Received2019-04-29
Model NumberUS910
Catalog NumberUS910
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerAESCULAP INC.
Manufacturer Address615 LAMBERT POINTE DRIVE HAZELWOOD MO 63042 US 63042


Patients

Patient NumberTreatmentOutcomeDate
10 2019-04-29