3M? CUROS JET? DISINFECTING CAP FOR NEEDLELESS CONNECTORS N/A CFJ1-270

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2019-04-30 for 3M? CUROS JET? DISINFECTING CAP FOR NEEDLELESS CONNECTORS N/A CFJ1-270 manufactured by 3m Health Care.

Event Text Entries

[143568934] Current product labeling notes the potential for intermittent leaking with baxter? Continu-flo? Solution set with clearlink? Luer activated valve. Please see the exact wording below: caution: there is a potential for intermittent leaking when baxter? Continu-flo? Solution set with clearlink? Luer activated valve or bd maxzero? Needle-free connectors are used under pressure with 3m? Curos jet? Disinfecting caps (cfj1-270 and cfj5-250. ) monitor these connectors if used under pressure. Alternatively, 3m? Curos? Disinfecting cap for needleless connectors (cff1-270 and cff10- 250) may be used.
Patient Sequence No: 1, Text Type: N, H10


[143568935] A nurse reported 3m? Curos jet? Disinfecting caps were placed on the needleless connector ports of a patient's iv tubing. The nurse reported a curos cap "failed" on one of the needleless connector ports and a puddle of propofol iv medication was found on the floor underneath the iv port. The nurse reported this resulted in the patient not receiving the appropriate dose of medication. No additional information has been received to date.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2110898-2019-00069
MDR Report Key8565291
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2019-04-30
Date of Report2019-04-30
Date Mfgr Received2019-04-02
Date Added to Maude2019-04-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactDIANNE GIBBS
Manufacturer StreetBUILDING 275 5W 06
Manufacturer CityST. PAUL MN 55144
Manufacturer CountryUS
Manufacturer Postal55144
Manufacturer Phone6517379117
Manufacturer G13M COMPANY 3M COLUMBIA
Manufacturer Street5400 PARIS ROAD
Manufacturer CityCOLUMBIA MO 65202
Manufacturer CountryUS
Manufacturer Postal Code65202
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand Name3M? CUROS JET? DISINFECTING CAP FOR NEEDLELESS CONNECTORS
Generic NameCUROS JET DISINFECTING CAP
Product CodeLKB
Date Received2019-04-30
Model NumberN/A
Catalog NumberCFJ1-270
Lot NumberUNKNOWN
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
Manufacturer3M HEALTH CARE
Manufacturer Address2510 CONWAY AVE ST. PAUL MN 55144 US 55144


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2019-04-30

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.