DISSOLUTION UNIT-TESTED G047-80101

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2019-04-30 for DISSOLUTION UNIT-TESTED G047-80101 manufactured by Concord Manufacturing.

Event Text Entries

[143618179] The plant investigation is in process. A supplemental mdr will be submitted upon completion of this activity.?
Patient Sequence No: 1, Text Type: N, H10


[143618180] A biomedical technician (biomed) at a user facility reported that a fresenius dissolution unit had a visibly burned fill valve. The burned component was found while servicing the machine for a report of smoke that was observed coming from the machine during use. A patient was not connected to the machine at the time of the incident and there was no harm to any patients or individuals because of this malfunction. The burned fill valve was the original fresenius part on the machine. The biomed reported that there was no damage observed on any other components, or any other additional issues, associated with the burned fill valve. The biomed replaced the fill valve, which resolved the issue. The unit was returned to service at the user facility without issue and without reoccurrence of the event. The fill valve was discarded by the biomed and is not available to be returned to the manufacturer for physical evaluation.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2937457-2019-01269
MDR Report Key8566845
Report SourceUSER FACILITY
Date Received2019-04-30
Date of Report2019-05-15
Date of Event2019-04-19
Date Mfgr Received2019-05-14
Device Manufacturer Date2008-07-02
Date Added to Maude2019-04-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMATTHEW AMARAL
Manufacturer Street920 WINTER ST.
Manufacturer CityWALTHAM MA 02451
Manufacturer CountryUS
Manufacturer Postal02451
Manufacturer Phone7816999758
Manufacturer G1CONCORD MANUFACTURING
Manufacturer StreetDIRECTOR, QUALITY SYSTEMS 4040 NELSON AVENUE
Manufacturer CityCONCORD CA 94520
Manufacturer CountryUS
Manufacturer Postal Code94520
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameDISSOLUTION UNIT-TESTED
Generic NameBLENDER/MIXER
Product CodeJRO
Date Received2019-04-30
Catalog NumberG047-80101
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeMO
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerCONCORD MANUFACTURING
Manufacturer Address4040 NELSON AVENUE CONCORD CA 94520 US 94520


Patients

Patient NumberTreatmentOutcomeDate
10 2019-04-30

© 2024 FDA.report