30-70ML CELL FREEZE CRYOGENIC STORAGE CONTAINER CF-250

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-05-01 for 30-70ML CELL FREEZE CRYOGENIC STORAGE CONTAINER CF-250 manufactured by Charter Medical, Ltd..

Event Text Entries

[143833579] The sample was not returned for investigation. There were no pictures or other information provided, which may have indicated an exact failure mode.
Patient Sequence No: 1, Text Type: N, H10


[143833580] The customer reported a hole under a port at the seam, which resulted in the product spilling out. The hole was noticed during the thawing process. The hole occurred in a laboratory setting and did not involve a patient. The product inside the cf-250 was exposed and was not infused into a patient. There was no report of an adverse event or need for additional medical intervention as a result of this occurrence. There was no report of laboratory/medical staff exposure as a result of this incident.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1066733-2019-00002
MDR Report Key8569018
Date Received2019-05-01
Date of Report2019-05-01
Date of Event2019-03-01
Date Mfgr Received2019-03-18
Device Manufacturer Date2019-01-03
Date Added to Maude2019-05-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. TODD MEINECKE
Manufacturer Street3948-A WESTPOINT BLVD.
Manufacturer CityWINSTON SALEM NC 27103
Manufacturer CountryUS
Manufacturer Postal27103
Manufacturer Phone3367686447
Manufacturer G1CHARTER MEDICAL, LTD.
Manufacturer Street3948-A WESTPOINT BLVD.
Manufacturer CityWINSTON SALEM NC 27103
Manufacturer CountryUS
Manufacturer Postal Code27103
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand Name30-70ML CELL FREEZE CRYOGENIC STORAGE CONTAINER
Generic Name30-70ML CELL FREEZE CRYOGENIC STORAGE CONTAINER
Product CodeLPZ
Date Received2019-05-01
Returned To Mfg2019-06-06
Model NumberCF-250
Catalog NumberCF-250
Lot Number151603
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCHARTER MEDICAL, LTD.
Manufacturer Address3948-A WESTPOINT BLVD. WINSTON SALEM NC 27103 US 27103


Patients

Patient NumberTreatmentOutcomeDate
10 2019-05-01