[143833596] It was reported that the plastic sleeve of the reprocessed ethicon's endopath xcel blunt tip trocar, w/smooth sleeve and adjustable plug, (purple) 12mm x 100mm, broke off inside the patient during an unidentified case. Per report, the broken piece was successfully retrieved from the patient through an unspecified method. There was no report of a serious injury, an adverse patient consequence or prolonged anesthesia related to this incident. Due to the reported event and required medical intervention to retrieve the broken piece from the surgical site, this medwatch is being filed. The device was returned for evaluation and the complaint was confirmed. A review of the reprocessing records was performed and indicated that all processes were conducted as required. A definitive root cause could not be determined at this time. No additional information is available. If additional information becomes available, a supplemental medwatch will be filed.
Patient Sequence No: 1, Text Type: N, H10


[143833777] It was reported that the plastic sleeve of the reprocessed ethicon trocar broke off inside the patient during an unidentified procedure.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3032391-2019-00009
MDR Report Key8570159
Date Received2019-05-01
Date of Report2019-05-01
Date of Event2019-04-09
Date Mfgr Received2019-04-09
Device Manufacturer Date2019-03-12
Date Added to Maude2019-05-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSTEPHEN WILSON
Manufacturer Street1500 NE HEMLOCK AVE
Manufacturer CityREDMOND OR 97756
Manufacturer CountryUS
Manufacturer Postal97756
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameETHICON
Generic NameENDOPATH? XCEL? BLUNT TIP TROCAR, W/SMOOTH SLEEVE & ADJUSTABLE PLUG, (PURPLE)
Product CodeNLM
Date Received2019-05-01
Returned To Mfg2019-04-24
Catalog NumberH12LP2RH
Lot Number413307
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMEDLINE RENEWAL
Manufacturer Address1500 NE HEMLOCK AVE REDMOND OR 97756 US 97756


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2019-05-01