ZRA

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a consumer report with the FDA on 2019-05-01 for ZRA manufactured by Tisport, Llc.

Event Text Entries

[143807983] We reviewed the dhr for this chair, and it passed all applicable quality tests and configuration requirements, and met specifications as ordered by the dealer when it left the facility. The dealer and/or a therapist measured the user for a wheelchair. Tisport as the manufacturer builds the chair based on the measurements provided to us from the dealer and/or therapist. All documentation and specifications for the chair were appropriately disseminated to the health specialist working with the end user. The device was not returned for evaluation. Based on the evidence provided, the fracture in the backrest bracket is likely the result of excessive forces introduced to the backrest assembly through impact or abuse. If more information is provided that changes the conclusions of the investigation, a follow up medwatch form 3500a will be filed.
Patient Sequence No: 1, Text Type: N, H10


[143807984] The end user alleges that while putting on his pants in the bathroom, the backrest hinges fractured, resulting in ligament damage in his arm. He is in a full arm cast, and will receive injections in his arm as treatment with the possibility of surgery for future treatment.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3032618-2019-00004
MDR Report Key8571403
Report SourceCONSUMER
Date Received2019-05-01
Date of Report2019-05-01
Date of Event2019-04-02
Date Mfgr Received2019-04-02
Device Manufacturer Date2015-06-24
Date Added to Maude2019-05-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. BRIAN ENGLISH
Manufacturer Street2701 W COURT ST
Manufacturer CityPASCO WA 99301
Manufacturer CountryUS
Manufacturer Postal99301
Manufacturer Phone5095866117
Single Use3
Remedial ActionRP
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameZRA
Generic NameWHEELCHAIR
Product CodeIOR
Date Received2019-05-01
Model NumberZRA
OperatorLAY USER/PATIENT
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerTISPORT, LLC
Manufacturer Address2701 W COURT ST PASCO WA 99301 US 99301


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2019-05-01

© 2026 FDA.report