VALVE IN CARTRIDGE, SPIRATION VALVE SYSTEM UNKNOWN

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-05-01 for VALVE IN CARTRIDGE, SPIRATION VALVE SYSTEM UNKNOWN manufactured by Spiration.

Event Text Entries

[143799590] Based on a letter to the editor in journal of bronchology & interventional pulmonology (an unusual complication of long-term endobronchial valves placed for persistent air leak, journal of bronchology & interventional pulmonology 24 (2): e18? 20), spiration became aware that two of three valves placed in a patient for a prolonged air leak post-lobectomy appeared to have migrated into the pleural space. According to the letter, the valves were placed under the hde and the air leak resolved, but the valves were not removed within 6 weeks as specified in hde instructions for use. When seen 2 years later by the physicians who submitted the letter, the valves were still in place and they report that the patient had recurring pseudomonas aeruginosa pneumonia requiring iv antibiotics for the two years following valve placement. Two of the three valves were removed and cultured, yielding pseudomonas aeruginosa. The authors indicate that while the infection is not necessarily due to the valves, it is possibly related to them. The third valve could not be retrieved. After removal of the two valves, the patient has done well without further antibiotics.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3004450998-2019-00005
MDR Report Key8571510
Date Received2019-05-01
Date of Report2019-05-01
Date Mfgr Received2019-04-03
Date Added to Maude2019-05-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA0
Report to FDA3
Event Location3
Manufacturer ContactCYNDY ADAMS
Manufacturer Street6675 185TH AVE NE
Manufacturer CityREDMOND WA 98052
Manufacturer CountryUS
Manufacturer Postal98052
Manufacturer Phone4256365450
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameVALVE IN CARTRIDGE, SPIRATION VALVE SYSTEM
Generic NameSPIRATION VALVE SYSTEM
Product CodeOAZ
Date Received2019-05-01
Model NumberUNKNOWN
Lot NumberUNKNOWN
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerSPIRATION
Manufacturer Address6675 185TH AVENUE NE REDMOND WA 98052 US 98052


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2019-05-01

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.