WEB SLS 27 SINGLE LAYER FG15090-001

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2019-05-01 for WEB SLS 27 SINGLE LAYER FG15090-001 manufactured by Sequent Medical, Inc.

Event Text Entries

[143805272] The lot number was provided. A review of the approved device history records indicated the lot met all release criteria. A lot history trending review was performed and there were no similar complaints for this lot number. The device was returned to the manufacturer for evaluation. The investigation is underway.
Patient Sequence No: 1, Text Type: N, H10


[143805273] It was reported that after delivery into the aneurysm, the web was protruding into the carotid artery. The web was retracted to achieve better positioning in the aneurysm; however, during the attempt, the delivery wire detached from the web. The carotid artery was almost 90% occluded by the proximal segment of the web. A snare device was used to capture the device marker and retrieve the web successfully. There was no reported patient injury. The patient was reported to be stable and was discharged the day after the procedure.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2032493-2019-00113
MDR Report Key8571727
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2019-05-01
Date of Report2019-04-02
Date of Event2019-03-27
Date Mfgr Received2019-04-02
Device Manufacturer Date2018-08-14
Date Added to Maude2019-05-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. DEBBY CALLAHAN
Manufacturer Street35 ENTERPRISE DRIVE
Manufacturer CityALISO VIEJO CA 92656
Manufacturer CountryUS
Manufacturer Postal92656
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameWEB SLS 27 SINGLE LAYER
Generic NameWOVEN ENDOBRIDGE (WEB) ANEURYSM EMBOLIZATION SYSTEM
Product CodeOPR
Date Received2019-05-01
Returned To Mfg2019-04-26
Model NumberFG15090-001
Lot Number18081414
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSEQUENT MEDICAL, INC
Manufacturer Address11 A COLUMBIA ALISO VIEJO CA 92656 US 92656


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2019-05-01

© 2026 FDA.report