ABL90 FLEX PLUS ANALYZER 393-092

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2019-05-02 for ABL90 FLEX PLUS ANALYZER 393-092 manufactured by Radiometer Medical Aps.

Event Text Entries

[146362106] According to the complaint received, a customer reported to have intermittent patient barcode labelling issues on an abl90 flex analyzer. The following cases were observed when different patient labels were scanned: only some of the digits in the medical record number (mrn) was detected and one of the digits was incorrect. The patient name was correct. The mnr is correctly detected. But the patient name is not detected. Based on the current investigation performed in r&d, the issue seems to be related to the patient demographic rather than the reading of the barcode. New data of the issue has been requested by radiometer to perform further investigation.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002807968-2019-00016
MDR Report Key8572625
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2019-05-02
Date of Report2019-06-14
Date of Event2019-03-22
Date Mfgr Received2019-05-20
Device Manufacturer Date2018-07-13
Date Added to Maude2019-05-02
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. BOLETTE DUUN-CHRISTENSEN
Manufacturer StreetAAKANDEVEJ 21
Manufacturer CityBROENSHOEJ, 2700
Manufacturer CountryEZ
Manufacturer Postal2700
Manufacturer G1RADIOMETER MEDICAL APS
Manufacturer StreetAAKANDEVEJ 21
Manufacturer CityBROENSHOEJ, 2700
Manufacturer CountryDA
Manufacturer Postal Code2700
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameABL90 FLEX PLUS ANALYZER
Generic NameABL90 FLEX PLUS ANALYZER
Product CodeMQM
Date Received2019-05-02
Model Number393-092
Catalog Number393-092
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerRADIOMETER MEDICAL APS
Manufacturer AddressAAKANDEVEJ 21 BROENSHOEJ, 2700 DA 2700


Patients

Patient NumberTreatmentOutcomeDate
10 2019-05-02

© 2026 FDA.report