MERGE UNITY PACS UNITY PACS R11.2.2

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2019-05-02 for MERGE UNITY PACS UNITY PACS R11.2.2 manufactured by Merge Healthcare.

Event Text Entries

[145617344] Merge unity pacs is designed to display a message to the user when an exam report fails to upload to the image server. In this instance, there was no evidence that the physician received the message warning that the report did not upload. It was not possible to determine if the message displayed on the screen for the physician when the report was first read and saved. The merge healthcare unity investigation showed the report did not upload successfully, and will have to be recreated. To resolve this issue, the exam was re-read by the doctor. No other review/investigation will be performed for cause.
Patient Sequence No: 1, Text Type: N, H10


[145617345] Merge unity pacs is a medical image and information management system that is used for viewing, selection, processing, printing, telecommunications, and media interchange of medical images from a variety of diagnostic imaging systems. On (b)(6) 2019 the customer reported an exam was read on (b)(6) 2019, however, the workstation had an operating system failure and the exam report could not be found. An investigation showed the report did not upload successfully. The report word template was created and the exam was re-read by the doctor on the same day (b)(6) 2019. There was no reported adverse event to the patient. However, while images are available, a finalized report not being available for subsequent review by the patient's physician could potentially result in a delay in care that could lead to harm. Reference complaint-(b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2183926-2019-00019
MDR Report Key8573091
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2019-05-02
Date of Report2019-04-02
Date of Event2019-04-02
Date Mfgr Received2019-04-02
Device Manufacturer Date2018-11-28
Date Added to Maude2019-05-02
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMEG MUCHA
Manufacturer Street900 WALNUT RIDGE DRIVE
Manufacturer CityHARTLAND WI 53029
Manufacturer CountryUS
Manufacturer Postal53029
Manufacturer G1MERGE HEALTHCARE
Manufacturer Street900 WALNUT RIDGE DRIVE
Manufacturer CityHARTLAND WI 53029
Manufacturer CountryUS
Manufacturer Postal Code53029
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMERGE UNITY PACS
Generic NamePICTURE AND ARCHIVING COMMUNICATIONS SYSTEM
Product CodeLLZ
Date Received2019-05-02
Model NumberUNITY PACS R11.2.2
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerMERGE HEALTHCARE
Manufacturer Address900 WALNUT RIDGE DRIVE HARTLAND WI 53029 US 53029


Patients

Patient NumberTreatmentOutcomeDate
10 2019-05-02

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.