MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-05-02 for PAXGENE? BLOOD CCFDNA TUBE 768165 manufactured by Becton, Dickinson And Company (bd).
[146510795]
Device evaluated by mfr: a device evaluation and/or device history review is anticipated, but is not complete. Upon completion, a supplemental report will be filed.
Patient Sequence No: 1, Text Type: N, H10
[146510796]
It was reported that paxgene? Blood ccfdna tube gave erroneous results. This occurred on 1000 occasions. The following information was provided by the initial reporter: unsatisfied quality. Due to poor performance compared to competitor's product will be returned. Paxgene ccfdna tube has poor quality on volume of ccfdna and it caused qc fail. ((b)(6)? S guarantee criteria: nipt 10ng/ul: 6 fail out of 17 samples, cancer 40n/ul). In library production step, paxgene tube was shown 37% lower performance in library concentration comparison. Sample condition: same day of collection? Yes. Same phlebotomist? Yes. Tube storage conditions (time and temp before centrifugation): centrifugation. Conditions for both tubes: 2900g / 10mins / 4degree. Tube lot numbers / expiration dates - lot# 8158948 / 31. July. 2019.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 9617032-2019-00534 |
MDR Report Key | 8573798 |
Date Received | 2019-05-02 |
Date of Report | 2019-06-27 |
Date of Event | 2019-04-17 |
Date Mfgr Received | 2019-04-18 |
Device Manufacturer Date | 2018-06-07 |
Date Added to Maude | 2019-05-02 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | OTHER HEALTH CARE PROFESSIONAL |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | BDX BRETT WILKO |
Manufacturer Street | 9450 SOUTH STATE STREET |
Manufacturer City | SANDY UT 84070 |
Manufacturer Country | US |
Manufacturer Postal | 84070 |
Manufacturer Phone | 8015652341 |
Manufacturer G1 | BECTON, DICKINSON AND COMPANY (BD) |
Manufacturer Street | BELLIVER WAY BELLIVER INDUSTRIAL ESTATE |
Manufacturer City | PLYMOUTH |
Manufacturer Country | US |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 0 |
Brand Name | PAXGENE? BLOOD CCFDNA TUBE |
Generic Name | DNA TUBE |
Product Code | PJE |
Date Received | 2019-05-02 |
Catalog Number | 768165 |
Lot Number | 8158948 |
Device Expiration Date | 2019-07-31 |
Device Availability | * |
Device Age | DA |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | BECTON, DICKINSON AND COMPANY (BD) |
Manufacturer Address | BELLIVER WAY BELLIVER INDUSTRIAL ESTATE PLYMOUTH US |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 1. Other | 2019-05-02 |