GLOVE SURGICAL PROTEXIS PI MICRO UNKNOWN

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-05-02 for GLOVE SURGICAL PROTEXIS PI MICRO UNKNOWN manufactured by Cardinal Health (thailand) 222 Limited.

Event Text Entries

[144139493] The actual catalog number (and lot number) of the protexis pi micro glove that was worn are not known. No other specifics about the event or the physician have been provided. The manufacturing facility has determined that the potential root cause is the use of an alternate new source for the polyisoprene which is used in the manufacture of the gloves. The new source of polyisoprene had less tear resistance than the previous source for the same thickness control. As of october 2018 cardinal health has reverted back to the original supplier and the first lot released in europe was ts18100005. The customer is returning "representative samples". However, the manufacturing facility does not need samples to be returned for evaluation as they identified the problem. If the "representative samples" are received and their evaluation reveals a different root cause, then a follow up report would be filed. We will continue to monitor complaints for such issues.
Patient Sequence No: 1, Text Type: N, H10


[144139494] During a cardiac procedure ((b)(6)), the protexis pi micro gloves tore, and the surgeon had blood on his hands. The patient was (b)(6). No other information has been provided.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1423537-2019-00314
MDR Report Key8573977
Date Received2019-05-02
Date of Report2019-05-02
Date of Event2019-04-08
Date Mfgr Received2019-04-08
Date Added to Maude2019-05-02
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactPATRICIA TUCKER
Manufacturer Street1500 WAUKEGAN RD
Manufacturer CityWAUKEGAN IL 60085
Manufacturer CountryUS
Manufacturer Postal60085
Manufacturer Phone8478874151
Manufacturer G1CARDINAL HEALTH (THAILAND) 222 LIMITED
Manufacturer Street7/111 MOO 4 HIGHWAY #331, MABY
Manufacturer CityRAYONG 21140
Manufacturer CountryTH
Manufacturer Postal Code21140
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameGLOVE SURGICAL PROTEXIS PI MICRO
Generic NameSURGEON'S GLOVES
Product CodeKGO
Date Received2019-05-02
Catalog NumberUNKNOWN
Lot NumberUNKNOWN
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerCARDINAL HEALTH (THAILAND) 222 LIMITED
Manufacturer Address7/111 MOO 4 HIGHWAY #331, MABY RAYONG 21140 TH 21140


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2019-05-02