VITROS SALI CHEMISTRY SLIDES 1314343

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2007-05-31 for VITROS SALI CHEMISTRY SLIDES 1314343 manufactured by Ortho-clinical Diagnostics.

Event Text Entries

[16722382] A customer observed repeatable negatively biased sali results during a correlation study on the 5, 1 fs analyzer. No biased results were reported. Biased results of the direction and magnitude observed may lead to inappropriate physician action. There was no report of pt harm as a result of this event. This report corresponds to ortho clinical diagnostics inc.
Patient Sequence No: 1, Text Type: D, B5


[16899791] Investigation into this event found that the results did not correlate with another oem method. Repeat testing results using an alternate slide cart matched the original results. Qc results were acceptable. Calibrator responses and calibration parameters were acceptable. After obtaining fresh samples, acceptable correlation was obtained between vitros and the oem method. The biased results are confined to the original samples, and the customer is investigating the transfer containers from the original samples. The root cause of the biased results has not been determined.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1319809-2007-00151
MDR Report Key857656
Report Source05
Date Received2007-05-31
Date of Report2007-05-03
Date of Event2007-05-03
Date Mfgr Received2007-05-03
Device Manufacturer Date2007-02-01
Date Added to Maude2008-02-04
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactTHOMAS SMITH
Manufacturer Street100 INDIGO CREEK DR
Manufacturer CityROCHESTER NY 146525101
Manufacturer CountryUS
Manufacturer Postal146525101
Manufacturer Phone5854533735
Manufacturer G1ORTHO-CLINICAL DIAGNOSTICS
Manufacturer Street300 WEILAND RD.
Manufacturer CityROCHESTER NY 14652
Manufacturer CountryUS
Manufacturer Postal Code14652
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVITROS SALI CHEMISTRY SLIDES
Generic NameIN VITRO DIAGNOSTIC
Product CodeDKJ
Date Received2007-05-31
Model NumberNA
Catalog Number1314343
Lot Number2226-0049-8138
ID NumberNA
Device Expiration Date2008-02-01
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Implant FlagN
Date RemovedB
Device Sequence No1
Device Event Key957640
ManufacturerORTHO-CLINICAL DIAGNOSTICS
Manufacturer Address100 INDIGO CREEK DR ROCHESTER NY 146525101 US


Patients

Patient NumberTreatmentOutcomeDate
10 2007-05-31

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.