EVOLUTION? COLONIC CONTROLLED-RELEASE STENT - UNCOVERED G48038 EVO-25-30-10-C

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2019-05-03 for EVOLUTION? COLONIC CONTROLLED-RELEASE STENT - UNCOVERED G48038 EVO-25-30-10-C manufactured by Cook Ireland Ltd.

Event Text Entries

[143949768] 510(k) number: k163468. (b)(4). Investigation is still pending, a follow up mdr will be submitted to include the investigation conclusions.
Patient Sequence No: 1, Text Type: N, H10


[143949769] User advanced the delivery system through wire guide to desired stenosis area. User placed the delivery system at 3-4cm front of stenosis area and ready to release the stent. User pulled the trigger each time and the sheath backed a little, and meanwhile the yellow marker on the stent always stays at 3-4 cm front of stenosis. User saw the tip of the stent open through x-ray, but the stent suddenly move forward and detached from sheath while user release the stent in half. The yellow marker on the tip of stent still remains at 3-4cm front of stenosis, user cannot retract the stent by pressing the retracting trigger. User saw the whole stent squeeze in 4cm long 4cm in front of stenosis area. User left the stent in patient colon and placed another stent at stenosis area with no issue.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3001845648-2019-00191
MDR Report Key8577446
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2019-05-03
Date of Report2019-05-31
Date of Event2019-03-04
Date Facility Aware2019-05-03
Date Mfgr Received2019-04-09
Device Manufacturer Date2018-11-09
Date Added to Maude2019-05-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. HEATHER RYAN
Manufacturer StreetO HALLORAN ROAD NATIONAL TECHNOLOGY PARK
Manufacturer CityLIMERICK
Manufacturer Phone061334440
Manufacturer G1COOK IRELAND LTD
Manufacturer StreetO HALLORAN ROAD NATIONAL TECHNOLOGY PARK
Manufacturer CityLIMERICK
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameEVOLUTION? COLONIC CONTROLLED-RELEASE STENT - UNCOVERED
Generic NameMQR STENT, COLONIC METALLIC EXPANDABLE
Product CodeMQR
Date Received2019-05-03
Returned To Mfg2019-05-01
Model NumberG48038
Catalog NumberEVO-25-30-10-C
Lot NumberC1556613
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Age115 DA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCOOK IRELAND LTD
Manufacturer AddressO HALLORAN ROAD LIMERICK


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2019-05-03