COBAS 6800 05524245001

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-05-03 for COBAS 6800 05524245001 manufactured by Roche Molecular Systems, Inc..

Event Text Entries

[146372568] For the customer's instrument, the local field service engineer (fse) went onsite, and found several positions of the processing transfer head front failing the tightness check as well as evidence of droplets/leakage. The issue has been solved with these service actions and the instrument is running without issues with passing tightness check results and with no further evidence of leakage. Corrective and preventive actions will be implemented as appropriate. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[146372569] The customer reported a high rate of invalid results with the cobas mpx test, and review of instrument files showed the generation of p07p flags, which is indicative of a volume error during supernatant removal. The local field service engineer (fse) went onsite, and found several positions of the processing transfer head front failing the tightness check as well as evidence of droplets/leakage. After replacing the o-rings & stop discs of these positions the check passed. The site has reported the invalid rate has decreased. There was no allegation of erroneous results, harm, or injury.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2243471-2019-00019
MDR Report Key8578567
Date Received2019-05-03
Date of Report2019-05-03
Date of Event2019-03-18
Date Mfgr Received2019-04-12
Date Added to Maude2019-05-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA STACIE-ANN CREIGHTON
Manufacturer Street1080 US HWY 202 S NA
Manufacturer CityBRANCHBURG NJ 08876
Manufacturer CountryUS
Manufacturer Postal08876
Manufacturer Phone9082537112
Manufacturer G1ROCHE INTERNATIONAL ROTKREUZ
Manufacturer StreetFORRENSTRASSE 2 NA
Manufacturer CityROTKREUZ 6343
Manufacturer CountryCH
Manufacturer Postal Code6343
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report0

Device Details

Brand NameCOBAS 6800
Generic NameAUTOMATED BLOODBORNE PATHOGEN TEST EQUIPMENT
Product CodeMZA
Date Received2019-05-03
Model NumberNA
Catalog Number05524245001
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device Availability*
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE MOLECULAR SYSTEMS, INC.
Manufacturer Address1080 US HIGHWAY 202 SOUTH NA BRANCHBURG NJ 08876 US 08876


Patients

Patient NumberTreatmentOutcomeDate
10 2019-05-03

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.