INVACARE MATTRESS RP8070M

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a consumer report with the FDA on 2019-05-06 for INVACARE MATTRESS RP8070M manufactured by Posey Products Llc.

Event Text Entries

[144160904] Customer confirmed will not be returning product. Without the device the reported issue could not be confirmed and possible causes of the failure could not be determined. Historical review of the complaint database did not find any other related event. However, this product is a serviceable item that had been in use for several years without any maintenance performed. It's possible the reported issue is related to normal wear and tear. At this time, there is no evidence that a manufacturing non-conformity contributed to the reported complaint and the instructions for use (ifu) were reviewed and determined to provide adequate instructions and warning for the safe and effective use of the device. Therefore, no corrective or preventative actions are necessary. All complaints are trended and reviewed by management on a monthly basis. As part of this monthly review, any excursion above the control limits for this failure mode will be assessed, documented and acted upon as warrant. Manufacturer reference file # (b)(4). Product will not be returning.
Patient Sequence No: 1, Text Type: N, H10


[144160905] Guardian reported the patient was found stuck between the mattress and the bed frame. It was noted the mattress was sunken in. It was confirmed the patient did not suffer any injuries from this event. The date the issue was discovered is unknown.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2020362-2019-00124
MDR Report Key8582896
Report SourceCONSUMER
Date Received2019-05-06
Date of Report2019-04-12
Date Mfgr Received2019-04-12
Date Added to Maude2019-05-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactWILLIAM HINCY
Manufacturer StreetPOSEY COMPANY 5635 PECK ROAD
Manufacturer CityARCADIA CA 91006
Manufacturer CountryUS
Manufacturer Postal91006
Manufacturer Phone6264433143
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameINVACARE MATTRESS
Generic NamePATIENT BED WITH CANOPY/RESTRAINTS
Product CodeOYS
Date Received2019-05-06
Model NumberRP8070M
Catalog NumberRP8070M
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerPOSEY PRODUCTS LLC
Manufacturer Address5635 PECK RD ARCADIA CA 91006 US 91006


Patients

Patient NumberTreatmentOutcomeDate
10 2019-05-06

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.