HMRS DISTAL FEMUR RIGHT 12OMM 63671112

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,other report with the FDA on 2019-05-06 for HMRS DISTAL FEMUR RIGHT 12OMM 63671112 manufactured by Stryker Orthopaedics-mahwah.

Event Text Entries

[144179987] Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies. There have been no other similar events for the lot referenced. It was noted that the device is not available for evaluation. Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.
Patient Sequence No: 1, Text Type: N, H10


[144179988] At (b)(6) 2008, hmrs placement of knee was performed by tumor resection. About 10 years after the surgery, the clip was pop out naturally from hmrs axis pin and pin was protruded from the femur component. Therefore, the revision surgery was done and the groove in which the clip fits was worn and could not fix the clip. Therefore, femur component, pin, clip and other attached parts were changed the new same devices. Dr 's request to check the wear of the clip groove in the femoral component.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number0002249697-2019-01857
MDR Report Key8583621
Report SourceHEALTH PROFESSIONAL,OTHER
Date Received2019-05-06
Date of Report2019-06-26
Date of Event2019-04-08
Date Mfgr Received2019-05-31
Device Manufacturer Date2008-01-30
Date Added to Maude2019-05-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. COLLIN NEITZEL
Manufacturer Street325 CORPORATE DRIVE
Manufacturer CityMAHWAH NJ 07430
Manufacturer CountryUS
Manufacturer Postal07430
Manufacturer Phone2018315000
Manufacturer G1STRYKER ORTHOPAEDICS-LIMERICK
Manufacturer StreetRAHEEN BUSINESS PARK
Manufacturer CityLIMERICK NA
Manufacturer Postal CodeNA
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameHMRS DISTAL FEMUR RIGHT 12OMM
Generic NameKNEE IMPLANT
Product CodeNPJ
Date Received2019-05-06
Catalog Number63671112
Lot NumberSFFEF1
Device Expiration Date2013-01-31
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerSTRYKER ORTHOPAEDICS-MAHWAH
Manufacturer Address325 CORPORATE DRIVE MAHWAH NJ 07430 US 07430


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization; 2. Required No Informationntervention 2019-05-06