MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a other,user facility report with the FDA on 2019-05-07 for BARD? NASOGASTRIC SUMP TUBE WITH PREVENT? ANTI-REFLUX FILTER 0046160 manufactured by C.r. Bard, Inc. (covington) -1018233.
[144268271]
The investigation is still in progress. Once the investigation is complete a supplemental report will be filed. The device was not returned.
Patient Sequence No: 1, Text Type: N, H10
[144268272]
It was reported that during the x-ray to confirm placement the ng tube could not be seen. The tube was placed in a patient with gastrointestinal bleeding and placement was confirmed by auscultation and from the return of gastric contents, but the x-ray did not demonstrate where the tube was placed. It was later reported per additional information from the complainant via email on (b)(6) 2019: the ng tube was continued to be used and was not replaced. There was not any reported impact to the patient.
Patient Sequence No: 1, Text Type: D, B5
| Report Number | 1018233-2019-02333 |
| MDR Report Key | 8585937 |
| Report Source | OTHER,USER FACILITY |
| Date Received | 2019-05-07 |
| Date of Report | 2019-06-05 |
| Date Mfgr Received | 2019-05-20 |
| Device Manufacturer Date | 2019-01-19 |
| Date Added to Maude | 2019-05-07 |
| Event Key | 0 |
| Report Source Code | Manufacturer report |
| Manufacturer Link | Y |
| Number of Patients in Event | 0 |
| Adverse Event Flag | 3 |
| Product Problem Flag | 3 |
| Reprocessed and Reused Flag | 3 |
| Health Professional | 3 |
| Initial Report to FDA | 3 |
| Report to FDA | 3 |
| Event Location | 3 |
| Manufacturer Contact | YONIC ANDERSON |
| Manufacturer Street | 8195 INDUSTRIAL BLVD |
| Manufacturer City | COVINGTON GA 30014 |
| Manufacturer Country | US |
| Manufacturer Postal | 30014 |
| Manufacturer Phone | 7707846100 |
| Manufacturer G1 | C.R. BARD, INC. (COVINGTON) -1018233 |
| Manufacturer Street | 8195 INDUSTRIAL BLVD |
| Manufacturer City | COVINGTON GA 30014 |
| Manufacturer Country | US |
| Manufacturer Postal Code | 30014 |
| Single Use | 3 |
| Previous Use Code | 3 |
| Event Type | 3 |
| Type of Report | 3 |
| Brand Name | BARD? NASOGASTRIC SUMP TUBE WITH PREVENT? ANTI-REFLUX FILTER |
| Generic Name | NG TUBE |
| Product Code | FEG |
| Date Received | 2019-05-07 |
| Model Number | 0046160 |
| Catalog Number | 0046160 |
| Lot Number | NGCZ2257 |
| Operator | HEALTH PROFESSIONAL |
| Device Availability | N |
| Device Age | DA |
| Device Eval'ed by Mfgr | R |
| Device Sequence No | 1 |
| Device Event Key | 0 |
| Manufacturer | C.R. BARD, INC. (COVINGTON) -1018233 |
| Manufacturer Address | 8195 INDUSTRIAL BLVD COVINGTON GA 30014 US 30014 |
| Patient Number | Treatment | Outcome | Date |
|---|---|---|---|
| 1 | 0 | 2019-05-07 |