HUDSON VENTILATOR TUBING SET,LONG LENGTH 1607

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2019-05-08 for HUDSON VENTILATOR TUBING SET,LONG LENGTH 1607 manufactured by Teleflex Medical.

Event Text Entries

[145633252] Qn#: (b)(4). One (1) 1607 ventilator tubing set, long length was received for investigation. The ventilator tubing set was visually inspected for any signs of abuse/misuse/damage. Nothing was noted. The dual limb 1607 ventilator tubing set, long length was setup for leak test on a lab leak tester. Per iso standard 5367 annex e, the ventilator tubing set was tested at 60 +/- 3 cmh2o of pressure with incoming equipment pressure set a 30 psi. The acceptable leak value is = 30 ml/min. The actual ml/min pressure recorded at test was 25 ml/min. The 1607 ventilator tubing set does leak, but well within the established iso acceptable parameters. The device history record of batch number 74l1802028 has been reviewed and no issues or discrepancies were found which could potentially relate to this complaint. No non-conformance reports were originated for the lot in question that can be associated to the complaint reported. The dhr shows that the product was assembled and inspected according to specifications. The ifu states for this product states, "ensure all connections are secure, all unused ports are capped, and water traps (if enclosed) are sealed properly". Based on the investigation performed, the reported complaint was confirmed. The complaint sample was tested against iso standard 5367 annex e which states at 60 +/- 3 cmh2o of pressure with incoming equipment pressure set a 30 psi the acceptable leak value is = 30 ml/min. The sample was well within the iso specified leakage value.
Patient Sequence No: 1, Text Type: N, H10


[145633253] The complaint is reported as: "product does not pass ventilator pressure test on npb980 vent - only reaches 72, not 75 minimum pressure". No patient involvement.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3004365956-2019-00104
MDR Report Key8590432
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2019-05-08
Date of Report2019-04-10
Date of Event2019-04-10
Date Mfgr Received2019-04-10
Device Manufacturer Date2018-11-15
Date Added to Maude2019-05-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactKATHARINE TARPLEY
Manufacturer Street3015 CARRINGTON MILL BLVD
Manufacturer CityMORRISVILLE NC 27560
Manufacturer CountryUS
Manufacturer Postal27560
Manufacturer Phone9194334854
Manufacturer G1TELEFLEX MEDICAL
Manufacturer StreetPARQUE INDUSTRIAL FINSA
Manufacturer CityNUEVO LAREDO 88275
Manufacturer CountryMX
Manufacturer Postal Code88275
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameHUDSON VENTILATOR TUBING SET,LONG LENGTH
Generic NameCIRCUIT, BREATHING (W CONNECTO
Product CodeCAI
Date Received2019-05-08
Returned To Mfg2019-04-17
Catalog Number1607
Lot Number74L1802028
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerTELEFLEX MEDICAL
Manufacturer AddressRESEARCH TRIANGLE PARK NC


Patients

Patient NumberTreatmentOutcomeDate
10 2019-05-08

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.