DISSOLUTION UNIT-TESTED DRY ACID MIXER- GRAN G047-80101

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,user f report with the FDA on 2019-05-08 for DISSOLUTION UNIT-TESTED DRY ACID MIXER- GRAN G047-80101 manufactured by Concord Manufacturing.

Event Text Entries

[144416235] The plant investigation is in process. A supplemental mdr will be submitted upon completion of this activity.?
Patient Sequence No: 1, Text Type: N, H10


[144416236] A biomedical technician (biomed) at a user facility reported that a fresenius dry acid mixer was seen smoking and giving off a burning smell during use. Upon inspection, the biomed found interior wiring that was visibly burned and melted. A patient was not connected to the machine at the time of the incident and there was no harm to any patients or individuals because of this malfunction. The wiring was the original fresenius part on the machine. The biomed reported that there was no damage observed on any other components, or any other additional issues, associated with the burned and melted wiring. The biomed confirmed that the dry acid mixer was being replaced rather than being repaired. The dry acid mixer was scheduled to be returned to the manufacturer for physical evaluation.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2937457-2019-01373
MDR Report Key8590912
Report SourceCOMPANY REPRESENTATIVE,USER F
Date Received2019-05-08
Date of Report2019-06-04
Date of Event2019-04-26
Date Mfgr Received2019-06-04
Device Manufacturer Date2003-04-22
Date Added to Maude2019-05-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMATTHEW AMARAL
Manufacturer Street920 WINTER ST.
Manufacturer CityWALTHAM MA 02451
Manufacturer CountryUS
Manufacturer Postal02451
Manufacturer Phone7816999758
Manufacturer G1CONCORD MANUFACTURING
Manufacturer StreetDIRECTOR, QUALITY SYSTEMS 4040 NELSON AVENUE
Manufacturer CityCONCORD CA 94520
Manufacturer CountryUS
Manufacturer Postal Code94520
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameDISSOLUTION UNIT-TESTED
Generic NameBLENDER/MIXER
Product CodeJRO
Date Received2019-05-08
Model NumberDRY ACID MIXER- GRAN
Catalog NumberG047-80101
Lot NumberN/A
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeMO
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerCONCORD MANUFACTURING
Manufacturer Address4040 NELSON AVENUE CONCORD CA 94520 US 94520


Patients

Patient NumberTreatmentOutcomeDate
10 2019-05-08

© 2026 FDA.report