MONITORING SYSTEM

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-05-07 for MONITORING SYSTEM manufactured by Vocera Communications, Inc..

Event Text Entries

[144677701] Pts on ge monitors and alarm notifications being routed to nursing staff via vocera. In this particular instance, two pts, being cared for by the same nurse, had critical desaturations within seconds of one another. The monitoring system generated alarm notifications for both pts. The notifications for each pt were delivered to the vocera badge of the correct staff member. However, since the alarm notifications were delivered to the vocera within seconds of one another, the vocera did not enunciate the second pt's notification or require a second acknowledgement. When the primary rn acknowledged the first pt's alarm notification, the second pt's alarm notification was automatically acknowledged, thus the staff member was unaware of the notification for the second pt. Fortunately for the second pt, a staff member in passing heard the alarm in the room and responded, thus no pt harm. This event was able to be recreated by pulling two code blue buttons at the same time; the staff members only had to acknowledge notification for the first alarm, and the second alarm notification was cancelled out. Vocera has responded and has been able to validate the issue. Fda safety report id# (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW5086453
MDR Report Key8591239
Date Received2019-05-07
Date of Report2019-05-03
Date of Event2019-04-22
Date Added to Maude2019-05-08
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationRISK MANAGER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Sequence Number: 1

Brand NameMONITORING SYSTEM
Generic NameSYSTEM, NETWORK AND COMMUNICATION, PHYSIOLOGICAL MONITOR
Product CodeMSX
Date Received2019-05-07
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrI
Device Sequence No1
Device Event Key0
ManufacturerVOCERA COMMUNICATIONS, INC.

Device Sequence Number: 2

Brand NameMONITORING SYSTEM
Generic NameSYSTEM, NETWORK AND COMMUNICATION, PHYSIOLOGICAL MONITOR
Product CodeMSX
Date Received2019-05-07
OperatorHEALTH PROFESSIONAL
Device Availability*
Device Eval'ed by MfgrI
Device Sequence No2
Device Event Key0
ManufacturerVOCERA COMMUNICATIONS, INC.


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2019-05-07