ASSAY TIP/CUP ELECSYS MODULARE170 12102137001

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2019-05-09 for ASSAY TIP/CUP ELECSYS MODULARE170 12102137001 manufactured by Roche Diagnostics.

Event Text Entries

[144555506] This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[144555507] The initial reporter complained of an issue with assay tip/cups from a particular batch number used with a cobas 6000 e 601 module with serial number (b)(4). There are cracks in the cups and the customer noticed a leak. When the customer switched to a different batch number there were no issues. A crack in the assay cup resulting in a leak could impact patient results. There was no allegation that questionable or erroneous results were generated at the customer site due to the leaking assay cup. There was no allegation that an adverse event occurred. The assay cups were requested for investigation.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2019-01740
MDR Report Key8594328
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2019-05-09
Date of Report2019-08-23
Date of Event2019-04-16
Date Mfgr Received2019-04-16
Date Added to Maude2019-05-09
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameASSAY TIP/CUP ELECSYS MODULARE170
Generic NameEQUIPMENT, LABORATORY, GENERAL PURPOSE
Product CodeLXG
Date Received2019-05-09
Model NumberNA
Catalog Number12102137001
Lot Number19817113
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2019-05-09

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.