ABL800 SERIES ABL825 393-825

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2007-05-24 for ABL800 SERIES ABL825 393-825 manufactured by Radiometer Medical Aps.

Event Text Entries

[662364] When measuring po2 on the blood gas analyzer with flexq, a physician considered that the obtained po2 values to be abnormally high. This was confirmed by new measurements. No alarm was issued. The values were close to the values of atmospheric air were obtained. There has been no information of serious injury in the reported information.
Patient Sequence No: 1, Text Type: D, B5


[7875040] The abnormally high po2 values are close to the values of atmospheric air which suggests, that air-bubbles has been introduced into the samples, and that the measurements may have been conducted on the secluded air.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3002807968-2007-00007
MDR Report Key859481
Report Source05,06
Date Received2007-05-24
Date of Report2007-05-24
Date of Event2007-04-23
Date Mfgr Received2007-04-24
Device Manufacturer Date2006-09-01
Date Added to Maude2007-07-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactGERT NIELSEN
Manufacturer StreetAKANDEVEJ 21
Manufacturer CityBRONSHOJ DK-2700
Manufacturer CountryDA
Manufacturer PostalDK-2700
Manufacturer Phone8273057
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameABL800 SERIES
Generic NameABL800FLEX
Product CodeCCE
Date Received2007-05-24
Model NumberABL825
Catalog Number393-825
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Implant FlagN
Date RemovedB
Device Sequence No1
Device Event Key855719
ManufacturerRADIOMETER MEDICAL APS
Manufacturer AddressBROENSHOEJ(COPENHAGEN DA DK-2700


Patients

Patient NumberTreatmentOutcomeDate
10 2007-05-24

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.