ANTERIOR CHAMBER CANNULA [RYCROFT] .80 X 22 MM (21GX7/8IN) 581281

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional report with the FDA on 2019-05-09 for ANTERIOR CHAMBER CANNULA [RYCROFT] .80 X 22 MM (21GX7/8IN) 581281 manufactured by Beaver-visitec International, Inc.

Event Text Entries

[144666131] No further patient injury information could be obtained by bvi. Bvi has concluded that this incident may be attributed to human error. No corrective action will be taken at this time. Test results from bvi indicates that no failure should occur during normal use when the cannula and syringe are properly installed based on the intended use of the product. Scientific literature is available regarding cannula-associated ocular injuries (coi), though incidences are rare, clinicians should be securely attached onto the connecting syringe.
Patient Sequence No: 1, Text Type: N, H10


[144666132] It was alleged the cannula, "flew off syringe into patient's eye while irrigation of the corneal wound was being performed at the end of the cataract surgery. A small tear occurred. "
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1211998-2019-00001
MDR Report Key8596926
Report SourceFOREIGN,HEALTH PROFESSIONAL
Date Received2019-05-09
Date of Report2019-05-08
Date of Event2019-04-02
Date Mfgr Received2019-04-02
Date Added to Maude2019-05-09
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. GENCI OMARI
Manufacturer Street500 TOTTEN POND RD 10 CITY POINT
Manufacturer CityWALTHAM MA 02451
Manufacturer CountryUS
Manufacturer Postal02451
Manufacturer Phone7819066925
Manufacturer G11211998
Manufacturer Street500 TOTTEN POND RD. 10 CITYPOINT
Manufacturer CityWALTHAM MA 02451
Manufacturer CountryUS
Manufacturer Postal Code02451
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameANTERIOR CHAMBER CANNULA [RYCROFT] .80 X 22 MM (21GX7/8IN)
Generic NameMANUAL OPHTHALMIC SURGICAL INSTRUMENT
Product CodeHMX
Date Received2019-05-09
Model NumberNA
Catalog Number581281
Lot Number6007422
Device AvailabilityN
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBEAVER-VISITEC INTERNATIONAL, INC
Manufacturer Address500 TOTTEN POND RD 10 CITYPOINT WALTHAM MA 02451 US 02451


Patients

Patient NumberTreatmentOutcomeDate
10 2019-05-09

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.