COOK CERVICAL RIPENING BALLOON W/STYLET G19891 J-CRBS-184000

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,other,study,user faci report with the FDA on 2019-05-10 for COOK CERVICAL RIPENING BALLOON W/STYLET G19891 J-CRBS-184000 manufactured by Cook Inc.

Event Text Entries

[144721427] As reported, the patient underwent placement of an optease vena cava filter. The indication for the filter placement was not reported. The filter subsequently malfunctioned and caused injury and damage to the patient, including but not limited to perforation and tilt. As a direct and proximate result of these malfunctions, the patient suffered life-threatening injuries and damages, and required extensive medical care and treatment. As a further proximate result, the patient has suffered and will continue to suffer significant medical expenses, pain and suffering, and other damages. The product was not returned for analysis and the sterile lot number has not been provided; therefore, no device analysis nor device history record review could be performed. The optease vena cava filter is indicated for use in the prevention of recurrent pulmonary embolism (pe) via percutaneous placement in the inferior vena cava (ivc) for patients in which anticoagulants are contraindicated, anticoagulant therapy for thromboembolic disease has failed, emergency treatment following massive pe where anticipated benefits of conventional therapy are reduced or for chronic, recurrent pe where anticoagulant therapy has failed, or is contraindicated. The purpose of a vena cava filter is to catch thrombus from the lower extremities as it travels along normal blood flow patterns up towards the heart. Without images or procedural films for review, the reported filter tilt and perforation events could not be confirmed and the exact cause could not be determined. Ivc filter tilt has been associated with the anatomy of the vessel, specifically asymmetry and tortuousness. Additionally, the timing and mechanism of the filter tilt is unknown. It is unknown if the tilt contributed to the reported perforation. A review of the instructions for use (ifu) notes vessel damage such as intimal tears and perforation as procedural complications related to ivc filters. Ivc perforation from removable filters is relatively common, and directly related to how long the filter has been in place. Studies have noted a greater than 80% perforation rate overall, with all filters imaged after 71 days from implantation revealing some level of perforation. Clinical factors that may have influenced the event include the patient? S pre-existing co-morbidities, pharmacological issues and lesion characteristics. Given the limited information available for review, there is nothing to suggest that a malfunction in the design and manufacturing process of the device; therefore, no corrective action will be taken. Should additional information become available, the file will be updated accordingly.
Patient Sequence No: 1, Text Type: N, H10


[144721428] It is reported in a study of methods of labor induction the patient was randomized to induction using a cook cervical ripening balloon w/stylet. When the cook cervical ripening balloon was removed, an artificial rupture of membranes was performed and meconium stained fluid was noted. The patient then experienced abnormal cardiotocogram tracing and failure to progress in labor. These events led to a category 2 emergency cesarean section delivery. The cesarean delivery and meconium stained fluid led to prolonged hospital stay. The provider (chief investigator) reported that there is no causal relationship between the cook cervical ripening balloon w/stylet and the meconium stained fluid, failure to progress in labor, abnormal cardiotocogram tracing, and resulting emergency cesarean delivery. No additional consequences have been reported as occurring. Additional details have been requested regarding the patient and event. At this time, no information has been provided.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1820334-2019-01047
MDR Report Key8600186
Report SourceFOREIGN,OTHER,STUDY,USER FACI
Date Received2019-05-10
Date of Report2019-05-24
Date of Event2019-01-26
Date Mfgr Received2019-05-23
Device Manufacturer Date2018-11-15
Date Added to Maude2019-05-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. LARRY POOL
Manufacturer Street750 DANIELS WAY
Manufacturer CityBLOOMINGTON IN 47404
Manufacturer CountryUS
Manufacturer Postal47404
Manufacturer Phone8123392235
Manufacturer G1COOK INC
Manufacturer Street750 DANIELS WAY
Manufacturer CityBLOOMINGTON IN 47404
Manufacturer CountryUS
Manufacturer Postal Code47404
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCOOK CERVICAL RIPENING BALLOON W/STYLET
Generic NamePFJ, DILATOR, CERVICAL, HYGROSCOPIC-IAMINARIA
Product CodePFJ
Date Received2019-05-10
Model NumberG19891
Catalog NumberJ-CRBS-184000
Lot Number9313833
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerCOOK INC
Manufacturer Address750 DANIELS WAY BLOOMINGTON IN 47404 US 47404


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization; 2. Other 2019-05-10

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.