EMERGIN INTELLISPACE EVENT MANAGEMENT VERSION 10.0 865354

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2019-05-13 for EMERGIN INTELLISPACE EVENT MANAGEMENT VERSION 10.0 865354 manufactured by Philips Medical Systems.

Event Text Entries

[144817822] Device evaluated by mfr: a follow up report will be submitted once the investigation is complete.
Patient Sequence No: 1, Text Type: N, H10


[144817823] The customer reported a serious injury where the customer alleged that the intellivue event management (iem) paging system failed to provide brady and asystole alarm alerts in a timely manner to the staff. The customer stated that from the piic ix to the central station to iem to the cisco phones it took 30 seconds. The customer stated that the iem orchestrator staff assignments escalation process sent the alerts to level1, then level2, then to level3 (charge nurse) where the customer alleged that it took too long to have the sites clinical team respond due to the escalation process delay. The customer alleged that as the escalation happened the patient was not attended to soon enough. The issue occurred on 01/01/2019 at 10:39:28am (in iem ml time) for room (b)(6). The customer stated that there was delay in treatment and the patient required resuscitation.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1218950-2019-03422
MDR Report Key8603316
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2019-05-13
Date of Report2019-05-06
Date of Event2019-01-16
Date Mfgr Received2019-05-06
Device Manufacturer Date2011-12-28
Date Added to Maude2019-05-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. BETTY HARRIS
Manufacturer Street3000 MINUTEMAN ROAD
Manufacturer CityANDOVER MA 01810
Manufacturer CountryUS
Manufacturer Postal01810
Manufacturer Phone9786871501
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameEMERGIN INTELLISPACE EVENT MANAGEMENT VERSION 10.0
Generic NameEMERGIN INTELLISPACE EVENT MANAGEMENT VERSION 10.0
Product CodeMSX
Date Received2019-05-13
Model Number865354
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerPHILIPS MEDICAL SYSTEMS
Manufacturer Address3000 MINUTEMAN ROAD ANDOVER MA 01810 US 01810


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2019-05-13

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.