ELECSYS CALCITONIN IMMUNOASSAY 06445853190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-05-14 for ELECSYS CALCITONIN IMMUNOASSAY 06445853190 manufactured by Roche Diagnostics.

Event Text Entries

[145224102] This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[145224103] The initial reporter complained of discrepant results for 1 patient sample tested for elecsys calcitonin immunoassay (calcitonin) on a cobas e 411 immunoassay analyzer. The initial result from the 411 analyzer was 3. 05 pg/ml. This result was reported outside of the laboratory. On (b)(6) 2019 the sample was repeated on the e411 analyzer with a result of 2. 89 pg/ml. The sample was also run twice by the elisa biomerica biochemmak method and the results were both > 200 pg/ml. The sample was sent to 2 other external laboratories where the result from the immulite method was "within the normal range" and the result from a different elisa method was > 200 pg/ml. There was no allegation that an adverse event occurred. The e411 analyzer serial number was (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2019-01787
MDR Report Key8607197
Date Received2019-05-14
Date of Report2019-06-26
Date of Event2019-04-17
Date Mfgr Received2019-04-23
Date Added to Maude2019-05-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report0

Device Details

Brand NameELECSYS CALCITONIN IMMUNOASSAY
Generic NameCALCITONIN TEST SYSTEM
Product CodeJKR
Date Received2019-05-14
Model NumberCALCITONIN
Catalog Number06445853190
Lot Number33757800
Device Expiration Date2019-10-31
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2019-05-14