MI-1000 061515-V

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor report with the FDA on 2019-05-14 for MI-1000 061515-V manufactured by Medical Illumination International Inc..

Event Text Entries

[145769101] The product from customer complaint # (b)(4) was received and an investigation was initiated. A visual inspection showed that the arm support fitting (p/n 1002036) failed with no other apparent damage. We performed a structural analysis iaw failure analysis of the mi-1000 dated 07/16/2015. We reviewed the porosity of the material and found nothing irregular. The product was rebuilt so we could perform capacity testing. The unit passed a 6x load test for 24+ hours. We were unable to determine the reason for the failure. The potential for misuse is always present but we were unable to definitively tie this failure to any abuse or misuse. This brakeage is unusual as this was the only reported brakeage of this type of the (b)(4) arms shipped in 2015. Note: this report is late because it was sent to the test emdr instead of production.
Patient Sequence No: 1, Text Type: N, H10


[145769102] On (b)(6) 2017 an incident was reported to medical illumination regarding an mi-1000 arm support fitting fractured causing the light to fall. No injuries reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2028295-2018-00001
MDR Report Key8609538
Report SourceDISTRIBUTOR
Date Received2019-05-14
Date of Report2018-01-23
Date of Event2017-10-25
Date Mfgr Received2017-10-25
Device Manufacturer Date2015-05-30
Date Added to Maude2019-05-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA0
Report to FDA3
Event Location3
Manufacturer ContactMR. ROBERT HESS
Manufacturer Street547 LIBRARY STREET
Manufacturer CitySAN FERNANDO CA 91340
Manufacturer CountryUS
Manufacturer Postal91340
Manufacturer Phone8188383025
Manufacturer G1MEDICAL ILLUMINATION INTRNATION INC.
Manufacturer Street547 LIBRARY STREET
Manufacturer CitySAN FERNANDO CA 91340
Manufacturer CountryUS
Manufacturer Postal Code91340
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMI-1000
Generic NameMINOR SURGICAL LIGHT
Product CodeFST
Date Received2019-05-14
Returned To Mfg2017-11-07
Model Number061515-V
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMEDICAL ILLUMINATION INTERNATIONAL INC.
Manufacturer Address547 LIBRARY STREET SAN FERNANDO CA 91340 US 91340