MI-1000 061514-V

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,distri report with the FDA on 2019-05-14 for MI-1000 061514-V manufactured by Medical Illumination Intrnation Inc..

Event Text Entries

[145768881] We received the returned light listed under customer complaint #(b)(4) and imitated an investigation. The product was a mi-1000 single ceiling mount. A visual inspection was performed and other than the arm support fitting (p/n 1002036) there was light surface damage found (ie: nicks and scratches). Because the arm support fitting (1002036) was tested several weeks ago under rma # (b)(4) for another customer complaint (#(b)(4)). The load test was not run again. We did cross-section analysis on the failed surfaces and did not see any porosity or abnormal material flow patterns. We could not definitively determine the cause of the failure. The light surface damage was deemed inconsequential. This are was out of (b)(4) sold in 2014 and this is the only failure of this component reported for that production year. Note: this report is late because it was sent to the test emdr instead of production.
Patient Sequence No: 1, Text Type: N, H10


[145768882] On (b)(6) 2017 an incident was reported to medical illumination regarding an mi-1000 arm support fitting fracturing causing the light to fall. There were no injuries.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2028295-2018-00002
MDR Report Key8609589
Report SourceCOMPANY REPRESENTATIVE,DISTRI
Date Received2019-05-14
Date of Report2018-01-25
Date of Event2017-11-30
Date Mfgr Received2017-11-30
Device Manufacturer Date2014-08-30
Date Added to Maude2019-05-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA0
Report to FDA3
Event Location3
Manufacturer ContactMR. ROBERT HESS
Manufacturer Street547 LIBRARY STREET
Manufacturer CitySAN FERNANDO CA 91340
Manufacturer CountryUS
Manufacturer Postal91340
Manufacturer Phone8188383025
Manufacturer G1MEDICAL ILLUMINATION INTRNATION INC.
Manufacturer Street547 LIBRARY STREET
Manufacturer CitySAN FERNANDO CA 91340
Manufacturer CountryUS
Manufacturer Postal Code91340
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMI-1000
Generic NameMINOR SURGICAL LIGHT
Product CodeFST
Date Received2019-05-14
Returned To Mfg2017-12-01
Model Number061514-V
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMEDICAL ILLUMINATION INTRNATION INC.
Manufacturer Address547 LIBRARY STREET SAN FERNANDO CA 91340 US 91340


Patients

Patient NumberTreatmentOutcomeDate
10 2019-05-14

© 2026 FDA.report