S5 GAS BLENDER SYSTEM 25-40-45

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2019-05-15 for S5 GAS BLENDER SYSTEM 25-40-45 manufactured by Livanova Deutschland.

Event Text Entries

[145106571] There was no patient involvement. Livanova (b)(4) manufactures the s5 gas blender system. The incident occurred in (b)(6). A livanova field service representative was dispatched to the facility to investigate. He tested the device and could reproduce the reported error message interrupting the connection to the gas supply. He re-connected the gas supply and the device was found to be working properly. The device has been requested back to the manufacturer site for further investigation. A review of the dhr could not identify any deviations or nonconformities relevant to the issue. If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
Patient Sequence No: 1, Text Type: N, H10


[145106572] Livanova (b)(4) received a report of s5 gas blender system displaying an error message during priming. There was no patient involvement.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9611109-2019-00348
MDR Report Key8611433
Report SourceHEALTH PROFESSIONAL
Date Received2019-05-15
Date of Report2019-07-23
Date Mfgr Received2019-06-26
Device Manufacturer Date2011-04-29
Date Added to Maude2019-05-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactBARBARA GALASSO
Manufacturer Street14401 W. 65TH WAY
Manufacturer CityARVAD CO 80004
Manufacturer CountryUS
Manufacturer Postal80004
Manufacturer G1LIVANOVA DEUTSCHLAND
Manufacturer StreetLINDBERGHSTR. 25
Manufacturer CityMUNICH 80939
Manufacturer CountryGM
Manufacturer Postal Code80939
Single Use3
Remedial ActionOT
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameS5 GAS BLENDER SYSTEM
Generic NameGAS CONTROL UNIT, CARDIOPULMONARY BYPASS
Product CodeDTX
Date Received2019-05-15
Model Number25-40-45
OperatorHEALTH PROFESSIONAL
Device Availability*
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerLIVANOVA DEUTSCHLAND
Manufacturer AddressLINDBERGHSTR. 25 MUNICH US


Patients

Patient NumberTreatmentOutcomeDate
10 2019-05-15

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.