PERCUTANEOUS PIGTAIL NEPHROSTOMY ACCESSORY SET G14095 080010

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2019-05-15 for PERCUTANEOUS PIGTAIL NEPHROSTOMY ACCESSORY SET G14095 080010 manufactured by Cook Inc.

Event Text Entries

[145511052] The investigation is in progress. A follow up report will be submitted should additional relevant information become available or upon completion of the investigation.
Patient Sequence No: 1, Text Type: N, H10


[145511053] It was initially reported on 19mar2019, the reporter is declaring a defect with the percutaneous pigtail nephrostomy set. During percutaneous placement of a drain in the renal pelvis of a urinoma the wire guide got stuck inside the dilator. Despite requests for additional patient and event details, no further information has been provided. On 15apr2019, during analysis of the returned 10fr pigtail catheter and. 038" teflon coated wire guide, it was observed that the guide is unraveled and the safety wire is broken. Based upon the condition of the returned device, this event has been determined to be a reportable malfunction.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1820334-2019-01165
MDR Report Key8612079
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2019-05-15
Date of Report2019-06-12
Date of Event2017-10-14
Date Mfgr Received2019-05-17
Device Manufacturer Date2017-02-13
Date Added to Maude2019-05-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. LARRY POOL
Manufacturer Street750 DANIELS WAY
Manufacturer CityBLOOMINGTON IN 47404
Manufacturer CountryUS
Manufacturer Postal47404
Manufacturer Phone8123392235
Manufacturer G1COOK INC
Manufacturer Street750 DANIELS WAY
Manufacturer CityBLOOMINGTON IN 47404
Manufacturer CountryUS
Manufacturer Postal Code47404
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePERCUTANEOUS PIGTAIL NEPHROSTOMY ACCESSORY SET
Generic NameKOB CATHETER, SUPRAPUBIC (AND ACCESSORIES)
Product CodeKOB
Date Received2019-05-15
Returned To Mfg2019-04-02
Model NumberG14095
Catalog Number080010
Lot Number7670729
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCOOK INC
Manufacturer Address750 DANIELS WAY BLOOMINGTON IN 47404 US 47404


Patients

Patient NumberTreatmentOutcomeDate
10 2019-05-15

© 2026 FDA.report