CAPITAL TEMP PUMP PROFESSIONAL TP700

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-05-15 for CAPITAL TEMP PUMP PROFESSIONAL TP700 manufactured by Stryker Medical-kalamazoo.

Event Text Entries

[145215198] The patient condition prior to the alleged incident was not disclosed, however, it was identified that there were no pre-existing conditions that would have made the patient susceptible to burns. The patient was immobile during her stay at the hospital and had to have assistance for all meals and movement. The patient had alleged shoulder pain prior to use of the t-pump and a physician ordered heat therapy to be applied but did not indicate the parameters of use. The t-pump was used with a mul-t-pad. The model and lot of the mul-t-pad were not recorded at the time of the event. The t-pump was alleged to be set at 100 degrees f (38 degrees c) and was used continuously with a 20 minute cycle for the night. The patient was put to bed between 17:30-18:30 on march 2nd and was then woken at 7:25 on march 3rd. There were no checks of the patients skin recorded during the time she was put to bed and the time she was awoken. The patient was allegedly diagnosed with a burn wound that was rinsed with sterile saline and had a foam dressing applied. The wound was noted to be closed and healed on march 19th. Information was provided to the user facility that the patient's skin should be checked regularly to ensure the condition is appropriate to continue therapy.
Patient Sequence No: 1, Text Type: N, H10


[145215199] A patient had superficial burns to her back and partial thickness burns to (2) areas with the worst over the left scapula. The patient had wound care provided following the event consisting of xerofoam to blisters with an optifoam gentle dressing.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number0001831750-2019-00566
MDR Report Key8613538
Date Received2019-05-15
Date of Report2019-05-15
Date of Event2019-03-03
Date Mfgr Received2019-04-17
Device Manufacturer Date2013-12-17
Date Added to Maude2019-05-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. BRIAN THOMPSON
Manufacturer Street3800 EAST CENTRE AVENUE
Manufacturer CityPORTAGE MI 49002
Manufacturer CountryUS
Manufacturer Postal49002
Manufacturer Phone2693292100
Manufacturer G1STRYKER MEDICAL-KALAMAZOO
Manufacturer Street3800 EAST CENTRE AVENUE
Manufacturer CityPORTAGE MI 49002
Manufacturer CountryUS
Manufacturer Postal Code49002
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameCAPITAL TEMP PUMP PROFESSIONAL
Generic NamePACK, HOT OR COLD, WATER CIRCULATING
Product CodeILO
Date Received2019-05-15
Catalog NumberTP700
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSTRYKER MEDICAL-KALAMAZOO
Manufacturer Address3800 EAST CENTRE AVENUE PORTAGE MI 49002 US 49002


Patients

Patient NumberTreatmentOutcomeDate
10 2019-05-15

© 2024 FDA.report