UL PRO FUSED HEADLIGHT CABLE 9FT, 275CM 001388LX9

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2019-05-16 for UL PRO FUSED HEADLIGHT CABLE 9FT, 275CM 001388LX9 manufactured by Integra York, Pa Inc..

Event Text Entries

[145429921] The product was returned for evaluation. A visual inspection found the connector is externally discolored consistent with impact from excessive heat. Note that the heat damage is consistent with use with the improperly functioning mlx light source. Also, the glass cap on the end of the connector is cracked and the internal fiber ends were observed to be melted. Further evaluation found, this cable has an internal separation at the "y" connection due to pull-force stress. This condition decreases light transmission. Functional testing of the cable found its ability to transmit light was extremely impaired. The root cause is considered user error, damage due to product misuse. No manufacturing, workmanship, or material deficiency has been identified. Device identifier: (b)(4). Linked to mfg # 2523190-2019-00058.
Patient Sequence No: 1, Text Type: N, H10


[145429922] It was reported that the 001388lx9, ul pro fused headlight cable 9 ft 275 cm, the external connectors have discolored, the glass cap on the connector cracked and internal melting was noted. In addition, the cable has separation at the "y" connection due to pull-force stress. It is unknown if there was any patient contact, injury or surgical delay. Request for additional information has been requested.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2523190-2019-00069
MDR Report Key8616984
Report SourceUSER FACILITY
Date Received2019-05-16
Date of Report2019-05-01
Date Mfgr Received2019-05-01
Date Added to Maude2019-05-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactUSER KIMBERLY SHELLY
Manufacturer Street311 ENTERPRISE DRIVE
Manufacturer CityPLAINSBORO NJ 08536
Manufacturer CountryUS
Manufacturer Postal08536
Manufacturer Phone6099362393
Manufacturer G1INTEGRA YORK, PA INC.
Manufacturer Street589 DAVIES DRIVE
Manufacturer CityYORK PA 17402
Manufacturer CountryUS
Manufacturer Postal Code17402
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameUL PRO FUSED HEADLIGHT CABLE 9FT, 275CM
Generic NameCABLES
Product CodeFST
Date Received2019-05-16
Returned To Mfg2019-04-16
Catalog Number001388LX9
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerINTEGRA YORK, PA INC.
Manufacturer Address589 DAVIES DRIVE 589 DAVIES DRIVE YORK PA 17402 US 17402


Patients

Patient NumberTreatmentOutcomeDate
10 2019-05-16

© 2026 FDA.report