MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2019-05-16 for FG GATEWAY OTW OUS 2.00MM X 15MM M0032072415200 manufactured by Boston Scientific - Minn.
        [145364715]
Subject device is not available.
 Patient Sequence No: 1, Text Type: N, H10
        [145364716]
It was reported that the subject balloon catheter was used during angioplasty of a 70% stenosis located in the middle cerebral artery (mca). The subject balloon catheter was pre-dilated in good condition; however, digital subtraction angiography (dsa) showed that the stenosis was quite hard; therefore, the physician continued dilating the subject balloon catheter to 8 atm when the balloon ruptured. Another balloon catheter was used to complete the procedure successfully. There were no clinical consequences reported due to this event.
 Patient Sequence No: 1, Text Type: D, B5
| Report Number | 3008853977-2019-00040 | 
| MDR Report Key | 8618380 | 
| Report Source | HEALTH PROFESSIONAL | 
| Date Received | 2019-05-16 | 
| Date of Report | 2019-07-11 | 
| Date of Event | 2019-04-25 | 
| Date Mfgr Received | 2019-06-24 | 
| Device Manufacturer Date | 2018-10-17 | 
| Date Added to Maude | 2019-05-16 | 
| Event Key | 0 | 
| Report Source Code | Manufacturer report | 
| Manufacturer Link | Y | 
| Number of Patients in Event | 0 | 
| Adverse Event Flag | 3 | 
| Product Problem Flag | 3 | 
| Reprocessed and Reused Flag | 3 | 
| Health Professional | 3 | 
| Initial Report to FDA | 3 | 
| Report to FDA | 3 | 
| Event Location | 3 | 
| Manufacturer Contact | MR. TARA LOPEZ | 
| Manufacturer Street | 47900 BAYSIDE PARKWAY | 
| Manufacturer City | FREMONT CA 94538 | 
| Manufacturer Country | US | 
| Manufacturer Postal | 94538 | 
| Manufacturer Phone | 5104132500 | 
| Manufacturer G1 | BOSTON SCIENTIFIC - MINN | 
| Manufacturer Street | ONE SCIMED PLACE | 
| Manufacturer City | MAPLE GROVE MN 55311 | 
| Manufacturer Country | US | 
| Manufacturer Postal Code | 55311 | 
| Single Use | 3 | 
| Previous Use Code | 3 | 
| Event Type | 3 | 
| Type of Report | 3 | 
| Brand Name | FG GATEWAY OTW OUS 2.00MM X 15MM | 
| Generic Name | CATHETER, BALLOON TYPE | 
| Product Code | GBA | 
| Date Received | 2019-05-16 | 
| Returned To Mfg | 2019-05-17 | 
| Catalog Number | M0032072415200 | 
| Lot Number | 22648046 | 
| Operator | HEALTH PROFESSIONAL | 
| Device Availability | R | 
| Device Age | DA | 
| Device Eval'ed by Mfgr | Y | 
| Device Sequence No | 1 | 
| Device Event Key | 0 | 
| Manufacturer | BOSTON SCIENTIFIC - MINN | 
| Manufacturer Address | ONE SCIMED PLACE MAPLE GROVE MN 55311 US 55311 | 
| Patient Number | Treatment | Outcome | Date | 
|---|---|---|---|
| 1 | 0 | 2019-05-16 |