MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,other report with the FDA on 2019-05-17 for FFRCT FFRCT 2.36.1.1 manufactured by Heartflow, Inc..
[145637895]
As part of heartflow's quality monitoring process, we identified a potential false negative and will be informing the ordering physician. Hfid # (b)(4): the investigation identified a potential false negative in the rca region; after the initial analysis was delivered, a second analysis completed as part of ongoing quality monitoring resulted in a decrease in the ffrct value from 0. 88 to 0. 62 on the distal rca. While heartflow has identified this issue, the physician has not provided feedback, nor indicated a safety event with the patient.
Patient Sequence No: 1, Text Type: N, H10
[145637896]
Heartflow identified a potential false negative result.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 3011276938-2019-00021 |
MDR Report Key | 8621618 |
Report Source | HEALTH PROFESSIONAL,OTHER |
Date Received | 2019-05-17 |
Date of Report | 2019-06-11 |
Date of Event | 2019-04-26 |
Date Mfgr Received | 2019-05-21 |
Device Manufacturer Date | 2019-04-26 |
Date Added to Maude | 2019-05-17 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MS. PATRICIA ERWIN |
Manufacturer Street | 1400 SEAPORT BLVD., BLDG. B |
Manufacturer City | REDWOOD CITY CA 940635594 |
Manufacturer Country | US |
Manufacturer Postal | 940635594 |
Manufacturer Phone | 6502414547 |
Manufacturer G1 | HEARTFLOW, INC. |
Manufacturer Street | 1400 SEAPORT BLVD., BLDG. B |
Manufacturer City | REDWOOD CITY CA 940635594 |
Manufacturer Country | US |
Manufacturer Postal Code | 940635594 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | FFRCT |
Generic Name | HEARTFLOW FFRCT |
Product Code | PJA |
Date Received | 2019-05-17 |
Model Number | FFRCT 2.36.1.1 |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Age | DA |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | HEARTFLOW, INC. |
Manufacturer Address | 1400 SEAPORT BLVD., BLDG. B REDWOOD CITY CA 940635594 US 940635594 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 1. Other | 2019-05-17 |