ZIPPIE VOYAGE EIZ16

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-05-17 for ZIPPIE VOYAGE EIZ16 manufactured by Sunrise Medical Privada Misiones.

Event Text Entries

[145484687] This is a known issue with the zippie voyage seating system. In investigating past incidents, sunrise medical's quality investigators and r&d engineers have found that there were several factors for such a failure as: improper mounting of the seating system by the care provider, where an audible click is heard when the seat is properly latched to the base. After market hardware used on the latching system which did not sit flush to the base, therefore causing the seat to not latch properly. Straps on the seat back system were not cleared out of the way when the seat was mounted on the base and obstructed the latch from connecting properly. Sunrise medical has made arrangements to have the stroller returned for a full evaluation. A follow up report may be filed to include any new relevant findings from the evaluation.
Patient Sequence No: 1, Text Type: N, H10


[145484688] The mother of an end user reported to sunrise medical via email an adverse event involving her son and a zippie voyage stroller. The mother claims the seat unlatched from the base while her son was in the seat and fell to the ground. She states he was taken to the hospital and claims that he received a "closed head injury" per the doctor's diagnosis. She stated the doctor didn't feel that a ct-scan was necessary and the end user was released.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9616084-2019-00005
MDR Report Key8621667
Date Received2019-05-17
Date of Report2019-04-22
Date of Event2019-04-22
Date Mfgr Received2019-04-22
Device Manufacturer Date2018-09-18
Date Added to Maude2019-05-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. GUSTAVO ZAMBRANO
Manufacturer Street2842 BUSINESS PARK AVE.
Manufacturer CityFRESNO CA 93727
Manufacturer CountryUS
Manufacturer Postal93727
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameZIPPIE VOYAGE
Generic NameSTROLLER (CHAIR WITH CASTERS)
Product CodeINM
Date Received2019-05-17
Model NumberEIZ16
OperatorLAY USER/PATIENT
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSUNRISE MEDICAL PRIVADA MISIONES
Manufacturer AddressNO. 110 PRIVADA INDUSTRIAL MISIONES DE LAS CALIFORNIAS TIJUANA, BAJA CALIFORNIA 22425 MX 22425


Patients

Patient NumberTreatmentOutcomeDate
101. Life Threatening 2019-05-17

© 2026 FDA.report